Download Print this page
Nidec PG-100 Instruction Manual

Nidec PG-100 Instruction Manual

Handy manometer

Advertisement

Quick Links

HANDY MANOMETER
PG-100
(Compliance with EMC Standards)
INSTRUCTION MANUAL  
1. HANDLING NOTES 
§
CAUTION
CAUTION
⑴The each attachment fitting torque to the pressure port should be less than 1 N・m.
⑵Availabie media to be measured is air or non corrosive gas only.
 The another media ,i.e. water or corrosive gas/liquid may damage the internal structure and cause malfunction.
⑶Applied pressure must be less than the specified maximum pressure.
⑷Internal battery
①Replace battery with Toshiba Battery Co.,Ltd./ERVM3.6V/ lithium battery only. Use of another battery may present a risk of fire or explosion. If you 
need replace battery, please place an order with us.
②When"LOBAT"sign is displayed, the battery should be replaced.
③The battery used in this device may present risk of fire of chemical burn if mistreated. Do not recharge, disassemble, heat above 100℃ (212℉) , or 
incinerate.
④Dispose of used battery promptly. Keep away from children. Do not disassemble and do not dispose of in fire.
⑸The only pH neutral detergents should be used to clean the body.
 The solvents such as thinners, benzine, etc. must be not used.
2. TRANSPORTATION & STORAGE 
§
⑴The product, which is a precision instrument, must be taken special care not to be damaged by impact nor by being dropped when it is trasported and in 
storage.
⑵The product must be avoided in storage where is dusty, dripping and vibrated.
Thank you very much for your purchase of our products.
This  instruction  manual  is  prepared  for  your  proper  use 
of the product.
Ver.3.1
This caution mark describes when there is a possibility that user may suffer from damage or physical 
damage may occur if the product is used improperly.
For more detailed information please ask for the nearest
distributor or the following sales center.
Nishi-Shinjuku Prime Square., 7-5-25
Nishi-Shinjuku Shinjuku-ku Tokyo 160-0023, Japan
https://www.nidec-copal-electronics.com
1

Advertisement

loading

Summary of Contents for Nidec PG-100

  • Page 1 HANDY MANOMETER Thank you very much for your purchase of our products. This  instruction  manual  is  prepared  for  your  proper  use  PG-100 of the product. For more detailed information please ask for the nearest distributor or the following sales center. (Compliance with EMC Standards) Nishi-Shinjuku Prime Square., 7-5-25 INSTRUCTION MANUAL   Nishi-Shinjuku Shinjuku-ku Tokyo 160-0023, Japan Ver.3.1 https://www.nidec-copal-electronics.com 1. HANDLING NOTES  § CAUTION This caution mark describes when there is a possibility that user may suffer from damage or physical  damage may occur if the product is used improperly. CAUTION ⑴The each attachment fitting torque to the pressure port should be less than 1 N・m. ⑵Availabie media to be measured is air or non corrosive gas only.  The another media ,i.e. water or corrosive gas/liquid may damage the internal structure and cause malfunction. ⑶Applied pressure must be less than the specified maximum pressure. ⑷Internal battery ①Replace battery with Toshiba Battery Co.,Ltd./ERVM3.6V/ lithium battery only. Use of another battery may present a risk of fire or explosion. If you  need replace battery, please place an order with us. ②When“LOBAT”sign is displayed, the battery should be replaced. ③The battery used in this device may present risk of fire of chemical burn if mistreated. Do not recharge, disassemble, heat above 100℃ (212℉) , or ...
  • Page 2 3. SPECIFICATION § G/V/R  ;  Gauge,  A  ;  Absolute ⑴  Indication of model P : kPa,  MP : MPa H : mmHg, B : bar No mark : kgf/cm (101G : mmH O) PG− 1 0 0 − ***   *   * Model  Name Series Series Pressure Range Max. Pressure Res.  Unit 101G 0〜1000 2000 101GP 0.00〜10.00 0.01 101RP −10.00〜10.00 0.01 102G 0〜1.000 0.001 kgf/cm 102GP...
  • Page 3 5. HOW TO USE  § ⑴ Set the attachment(Half Union or Barb Joint)into the pressure port. ⑵ Turn on the power. The display must be less than ± 5 counts.  (No pressure applied)   Adjust the zero point by Zero Adjustment  Trimmer with a minus driver to have“000”on the Pressure Display. !Do not adjust the zero point of 102AP/102AH because it is calibrated to a vacuum. ⑶ Set a tube into the attachment  Available tube size and material are as follows;For Half Union ;φ 6mm O.D.in Nylon or Urethane                         For Barb Joint ;φ 6mm O.D.,φ 4mm I.D.in Nylon ⑷ Apply the pressure to be measured. The value is displayed while the Hold Button is pressed, the value is kept. ⑸ The power can be turned off by pressing the Power Button or by keeping the power ON  over three minutes.  The latter is an additional function “auto power off” 6. WARRANTY  § Nidec Copal Electronics warrants the products for the period of one year after the date of the customerʼs receipt. We will repair the troubled products caused by  our improper designing and/or production control at our cost. Our warranty is limited to the products only, not on another damage that is caused by the product  malfunction. Please note that the repairing cost resulted from the following matters are out of our responsibility.  ⑴Trouble and damage caused by mishandling or careless usage against the handlimg manual.  ⑵Trouble and damage caused by improper remodeling, adjustment or repair.  ⑶Trouble and damage caused by natural disaster, fire or any other irresistible force.  ⑷Replacement of attachments. (Half Union and Barb Joint) .