Function Lock; Supervisory Mode; Available Pressure Ranges - Fluke 2700G Series User Manual

Reference pressure gauge
Hide thumbs Also See for 2700G Series:
Table of Contents

Advertisement

2700G Series
Users Manual

Function Lock

When set, access to each of the settable parameters
above can be turned "off" to prevent unauthorized
configuration changes. This is done with password
protection in Supervisory Mode. Push  to access
Supervisory Mode or  to go back to normal operation.

Supervisory Mode

If necessary, each user-configurable parameter can be
edited when you receive the Product. Some parameters
are locked and must be unlocked to configure them. Use
Supervisory mode to do this.
When you are in the Configure menu, and FUnC LOCK is
shown, it means that there are locked parameters.
To disable function lock:
1.
Push  . 0 PWRD is then shown.
2.
The password "101" is required to unlock
Supervisory mode. Push  to put in the password
entry. Hold  or  down to move faster through the
selections by a factor of 10. When you stop the
counter, push  and  again to move forward or
backward by a factor of 1. The password is factory
set and cannot be changed.
3.
Push .
From this point each parameter can be locked or
unlocked. Push  and  to select UnLOC or LOC for
8
sales@GlobalTestSupply.com
each parameter. To move to the subsequent parameter,
push .
You can access, lock, or unlock these functions:
Zero function (enable/disable)
Set pressure units (enable/disable)
Auto shutdown adjustment (enable/disable)
Damping settings (enable/disable)
Sample rate setting (enable/disable)
Tare setting (enable/disable)
Custom engineering units (set scale factor)
When a function is locked, it cannot be accessed or
changed from its current condition until you go to
Supervisory Mode and unlock it.

Available Pressure Ranges

Available pressure ranges are shown in the Specifications
section.