Using The Carescape Respiratory Modules And Compact Airway; Modules For O₂ Measurement; Selecting The O 2 Scale; Selecting The O 2 Sweep Speed - GE CARESCAPE User Manual

Modular monitors
Hide thumbs Also See for CARESCAPE:
Table of Contents

Advertisement

Airway gases
Using the CARESCAPE respiratory modules and
compact airway modules for O₂ measurement
Selecting the O
If the difference between FiO
1. Select a gas related parameter window.
2. Select the O2 tab > Setup.
3. Select an option from the Scale list.
Selecting the O
This selection affects the waveform.
1. Select a gas related parameter window.
2. Select the O2 tab > Setup.
3. Select an option from the O2 Sweep Speed list. The options are 0.625 mm/s,
6.25 mm/s, 12.5 mm/s, and 50 mm/s.
The smaller the value, the slower the sweep speed.
Setting O
alarms
2
1. Select a gas related parameter window.
2. Select the O2 tab > Alarms.
3. Check that the required alarm (EtO2 or FiO2) is on, and set its high and/or low
limit values.
Using the CARESCAPE respiratory modules
and compact airway modules for AA and N₂O
measurement

Selecting the agent scale

Every anesthetic agent has its own default scale that the monitor uses when detecting
the agent. You can change the scale of an agent if the amount used is higher than
the default scale. Default scales are given in the CARESCAPE Modular Monitors
Supplemental Information Manual.
1. Select a gas related parameter window.
2. Select the Agent/N2O tab > Setup.
3. Select an option from the Agent Scale list.

Selecting the agent sweep speed

This selection affects the waveform.
294
scale
2
and EtO
2
sweep speed
2
CARESCAPE Modular Monitors
is more than 6%, change the O
2
scale.