Invasive pressures
Do not turn on the IABP algorithm unless a balloon pump is in use.
●
If two pressure measurement modules that map to the same pressure channel
●
are connected, the first detected pressure measurement module is assigned to
the indicated pressure channel.
A pressure channel is activated when a pressure transducer interface cable is
●
connected to the PDM (Masimo), PDM (Nellcor), Tram 451, Tram 451N5, Tram
451M, Dual BP, BP/CO, BP, or BP/Dual Temp, or E-COP module, or when a pressure
transducer is connected to the E-COPSv, E-PSMP, E-PP, E-P, or E-PT module.
A pressure channel is deactivated when the pressure transducer interface cable is
●
disconnected from the PDM (Masimo), PDM (Nellcor), Tram 451, Tram 451N5, Tram
451M, Dual BP, BP/CO, BP, or BP/Dual Temp, or E-COP module, or when the pressure
transducer is disconnected from the or E-COPSv, E-PSMP, E-PP, E-P, or E-PT module.
E-modules: A deactivated pressure channel does not release the assigned pressure
●
channel. To release an assigned pressure channel, remove the module from the
system.
PCWP: Follow your care unit's policy and procedures for obtaining PCWP
●
measurements, including balloon inflation duration.
SPV and PPV: Measurement is reliable only when the patient is mechanically
●
ventilated.
Invasive pressure measurement setup
Invasive pressure equipment to patient connection
1. Module with invasive pressure measurement capability
2. Fluid bag with pressure infusor
3. Transducer setup
4. Invasive blood pressure adapter cable; single or dual cable (optional)
Invasive pressure module keys
There are invasive pressure keys on the following modules: