Using Alkaline Dry Cells - Honeywell Impact Series Operating Instructions And Maintenance Manual

Table of Contents

Advertisement

from the Off-Line Charger. Remove the instrument from the Base Station
and insert the Battery Packs. Tighten the retaining screw using the special
tool.
5.
The instrument is now ready for use.
Note:
The instrument does not have to be switched on during this procedure,
however if it is switched on a Warning 105 (see Appendix A) will be
shown. This should be accepted. In addition Fault 23 may be reported
when the instrument is removed from the Base Station. If this occurs,
the instrument should be switched off and back on again to reset the
instrument.

Using Alkaline Dry Cells

If using Alkaline Cells (Duracell
replacing, ensure that the batteries are not left out of the instrument for more than
15 minutes. This is to ensure correct biasing of the electrochemical CO
WARNING
The following points should be noted.
1.
The Bias PCB supplied with your electrochemical CO
rechargeable lithium battery. Do not attempt to recharge the battery under
any circumstances.
2.
Do not reuse the Bias PCB.
3.
Do not destroy the Bias PCB in fire.
4.
Dispose of the Bias PCB and cartridge packaging in accordance with Local
Regulations.
Impact / Impact Pro / Impact Pro IR Operating Instructions
®
MN1500 or Energizer
®
E91) and they require
sensor.
2
sensor contains a non-
2
63