9.
The sensors fitted to Impact Pro may give responses to gases other then
the target gas. The table below gives typical sensor responses to various
gases and should be used as a guide only. Note: Always use the target
gas to perform sensor calibrations.
10.
Your attention is drawn to the cross-sensitivity of the NH
presence of H
the NH
3
Applied
Gas
Sensor
CO
100 ppm
(100 ppm)
H
S
2
(40 ppm)
CO
2
(0.5% v/v)
Cl
(5 ppm)
2
NH
3
(50 ppm)
Hydrogen
200 ppm
(1000 ppm)
SO
2
(10 ppm)
1
Reading displayed as -0ppm.
** Electrochemical CO
58
S. This is normal operation and is due to the chemistry of
2
sensor.
Effect
Effect
on CO
on H
S
2
Sensor
0 ppm
0 ppm
40 ppm
0 ppm
0 ppm
0 ppm
0 ppm
0 ppm
0 ppm
0 ppm
0 ppm
0 ppm
cell only
2
Impact / Impact Pro / Impact Pro IR Operating Instructions
Effect
Effect
on CO
on Cl
2
2
Sensor**
Sensor
0% v/v
0 ppm
0% v/v
-25 ppm
0.5% v/v
0 ppm
0% v/v
5 ppm
0% v/v
-2 ppm
0% v/v
TBA
0% v/v
2 ppm
sensor in the
3
Effect
Effect
on NH
on SO
3
Sensor
Sensor
35 ppm
1 ppm
50 ppm
1 ppm
1
0 ppm
0 ppm
-2 ppm
50 ppm
1
35 ppm
0 ppm
10 ppm
2
TBA
1
TBA
TBA