Section 6

6.1 Calibrating the Temperature Sensor

Note: Measure the sample
temperature with a high precision
thermometer.

6.1.1 Temperature Conversion Table

Calibrating the Sensors
The free chlorine probe contains a temperature sensor. The sensor is factory
pre-set but requires on-site calibration in the sample.
1. From the Main screen, press the
2. Use the
LEFT ARROW
3. Use the
LEFT ARROW
4. Press
to select EXECUTION.
ENTER
5. Highlight EXECUTION. Press
Press
when the temperature stabilizes.
ENTER
6. Press the
RIGHT ARROW
the bottom left of the screen.
7. Adjust the displayed temperature to the actual value of the sample in degrees
Celsius. Press
. The deviation between the calibration and the sensor
ESC
theoretical response curve is provided.
Conversion from: °C to °F:
Conversion from °C to K:
Table 11 Temperature Conversions
°C
°F
0
32
1
33.8
2
35.6
3
37.4
4
39.2
5
41
6
42.8
7
44.6
8
46.4
9
48.2
10
50
11
51.8
12
53.6
13
55.4
14
57.2
15
59
16
60.8
17
62.6
18
64.4
RIGHT ARROW
key to highlight CALIBRATION. Press
key to highlight TEMP. CALIB. Press
. Wait for the temperature to appear.
ENTER
key to highlight the temperature values displayed on
×
°F
=
1.8 °C
+
32
K
=
°C
+
273.15
K
°C
273.15
23
274.15
24
275.15
25
276.15
26
277.15
27
278.15
28
279.15
29
280.15
30
281.15
31
282.15
32
283.15
33
284.15
34
285.15
35
286.15
36
287.15
37
288.15
38
289.15
39
290.15
40
291.15
41
key.
.
ENTER
.
ENTER
°F
K
73.4
296.15
75.2
297.15
77
298.15
78.8
299.15
80.6
300.15
82.4
301.15
84.2
302.15
86
303.15
87.8
304.15
89.6
305.15
91.4
306.15
93.2
307.15
95
308.15
96.8
309.15
98.6
310.15
100.4
311.15
102.2
312.15
104
313.15
105.8
314.15
33

Advertisement

Table of Contents
loading

This manual is also suitable for:

621-691-086

Table of Contents