Outputs 1/2
AFFECT
Conc. µA °C/°F
TYPE
0/20 or 4/20
MODE
Lin or Bi-lin
LOW
XXXX
MIDDLE
XXXX
UP
XXXX
Special Program
MODE
Preset/Last/Live
VALUE
XX
TEST
Used to test analog outputs in 1 mA steps (0–21 mA)
Figure 20
Linear and Bi-linear Output Loops

5.5 Setting Up the Display

Note: The factory-defined
programming language is English.
The instrument will revert to English
after a software upgrade or when
the default values are restored.
Table 7 mA Outputs Programming Parameters
Used to determine whether the analog output is assigned to current, concentration, or
temperature measurement.
Used to select the analog output range.
Choose a linear or bi-linear scale (see
Scale starting value
Scale middle value (bi-linear)
Scale ending value
Action of the analog output during calibration, system alarm, maintenance, or active
timer cycles:
Return to a preset value
Return to the last value before the event
Live = live measurement
Used to define the return value (0–21 mA)
20
(mA)
Bi-linear
4
0
10
20
1. From the Main display, press the
2. Use the
LEFT ARROW
3. Use the
LEFT ARROW
4. Program the parameters using
Figure
20).
40
50
60
30
70
(%)
RIGHT ARROW
key to highlight SERVICE. Press
key to highlight DISPLAY. Press
Table
8.
Programming the Instrument
Linear
80
90 100
key.
.
ENTER
.
ENTER
29