Fraction variant
New fraction configurations are made available in the new firmware. The following step must be performed for fraction
configured flowmeters:
①
Go to menu item 2.2.7.3 "Active Fraction" and press
and press
.
Note: If the ordered fraction is not found on the list, then select the last option "Empty".
Reinstate display and display lid
①
Reconnect display cable and carefully push display back into housing.
②
Remove the O-ring from the lid.
③
Reinstate the display lid until mechanical stop. Wind back the lid by one turn.
④
Mount the O-ring by pulling it over the display lid; then turn the lid until you feel friction from the O-ring on both sides.
⑤
Wind the display lid by one quarter of a turn to seal on the O-ring and tighten the fixing screw.
Siemens AG
Division Process Industries and Drives
Postfach 48 48
90026 NÜRNBERG
GERMANY
FC430 upgrade instructions - Compact
A5E33310535, 06/2016
8
. From the list, select the fraction used in the application,
FC430 upgrade instructions - Compact
A5E33310535-AG, 06/2016