11 Press "Enter" to save the entry.
12 Remove the conductivity sensor from the
top clip.
13 Screw the conductivity sensor into the cell
body cover and tighten the clamp connec-
tion.
14 Open the sample water inlet and outlet
again.
15 The conductivity electrode has its own
integrated temperature sensor. Variations
in the conductivity temperature measure-
ment can be adjusted via the menu
Tem.Offset.
This concludes the conductivity calibration.
6.7

Messages, alarms and errors

Messages, alarms and errors are displayed on the electronics module with the colored
sage symbol. Error messages can occur that can be acknowledged or that can not be acknowled-
ged. If several messages occur at the same time, the number of messages appears next to the
symbol. Press the
message symbol to display the message window. Configured alarms, mes-
sages that can be acknowledged and errors are displayed as clear text. A time-stamp shows
when the message was activated.
6.7.1 Acknowledgeable messages
They are acknowledged via the message window and the green ACK button
Error message
Maximum dosing
time?
Auto tune error
6.7.2 Non-acknowledgeable messages and errors
Error messages can only be rectified by eliminating the cause.
Error message
Measured value
display flashes
DI 1 flashes
DI 1
NOTICE
Cause
The maximum dosing time set for
a control output has been excee-
ded.
Auto tune terminated with error.
NOTICE
Cause
Measured value is outside the
measurement range.
Sample water flow rate recently
insufficient (delay time running).
Sample water flow rate insuffi-
cient for some time (delay time
elapsed).
6.6.5 Temperature calibration
1
Press the
symbol.
2
Press the
Calibration menu.
3
Tap the measurement "Temperature."
4
Press the "Cal. value" parameter.
5
Perform comparative temperature mea-
surement.
6
Use the keypad to enter the value to be
calibrated.
7
Press "Enter" to save the entry.
This concludes the temperature calibration.
Identify the cause, e.g. chemicals
tank empty.
Check the dosing pump.
Chapter 3.5.6.
Check measurement range and
change, if necessary. Check dosing
or controller settings.
Messwasser-Durchfluss prüfen
(ca. 33 l/h).
Schmutzfänger reinigen oder
wechseln.
Multi-Sensor falsch angeschlos-
sen oder defekt.
6. Operation
mes-
ACK
.
Remedy
Remedy
67