Schiller AT-101 Service Handbook page 62

12-channel ecg device
Hide thumbs Also See for AT-101:
Table of Contents

Advertisement

8
Appendix
8.1
Checklist
Reference/Testpoints
3.
The mains indicator lamp (6) switches off and the symbol battery is dis-
played on the LCD (7).
5.2.7 Paper Feed page 36
1.
The paper has to stop exactly at the perforation. If this is not the case, con-
trol the paper mark comparison.
5.2.8 Printer quality test page 36
1.
Check the blackness for regularity and good readability on the complete
print width.
2.
Check on the 25 mm/s printout if the space between 2 R peaks according
to the paper grids. The space should be 25 mm and may not show a dif-
ference of more than ± 0.5 mm.
5.2.9 Printing Speed page 36
1.
Check on the 25 mm/s printout if the space between 2 R peaks according
to the paper grids. The space should be 25 mm and may not show a dif-
ference of more than ± 0.5 mm.
5.2.10 Parallelism test page 37
1.
All calibration impulses should be lined up vertically and exactly below
one another.
2.
The maximum deviation may not be more than ± 0.5 square. If the values
are outside this tolerance, the mechanical adjustment of the print head
has to be corrected.
5.2.11 ECG amplifier page 37
1.
Check the amplitude height of the calibration impulse according to the pa-
per grid. It should be 20 mm ±0.5 square.
5.2.12 ECG lead and patient cable test page 38
1.
all leads (1) are blinking
2.
Uel for all leads is -350 to -550 mV
3.
all leads stops blinking
4.
Uel for all leads is -15 ...+15 mV
5.
Udif = 3985 ...4015 mV
6.
TPH = 18 ...30 °C
7.
Check amplitude and polarity according printouts
5.2.13 External printer test page 41
1.
all leads (1) are blinking
2.
Uel for all leads is -350 to -550 mV
3.
all leads stops blinking
4.
Uel for all leads is -15 ...+15 mV
5.
Udif = 3985 ...4015 mV
6.
TPH = 18 ...30 °C
7.
Check amplitude and polarity according printouts
5.2.14 Communication (RS-232) Test page 41
1.
The screen shows hex characters when the serial port is ok.
5.3 Safety Tests page 42
1.
The safety test is carried out in accordance with the EN 60601-1, Clause
18 and 19. This test may only be carried out with a tester that fulfils the
above mentioned norms and has been calibrated in accordance with ISO
norms.
Page 60
OK
False Remark
HR is set to 60
Add protocol to the checklist
A"#$0$