Assembly instructions
Assembling the correction
instrument
2
The correction instrument sleeve
CS 3361-05 (Sleeve correction
instrument, length 410 mm) and
the locking rod for the correction
instrument CS 3361-06 (Locking
rod correction instrument, length
410 mm) are stored in a pre-
assembled condition. To assemble
the instrument, first take the
locking rod out of the sleeve.
Set CS 3361-09 (Adjustor
correction instrument) on the
sleeve of CS 3361-01/-05
(Sleeve correction instrument,
length 300 mm/410 mm) and
tighten it counterclockwise by
means of the thread (left-handed
thread). It is important to firmly
tighten CS 3361-09 (Adjustor
correction instrument).
Then set CS 3361-02/-06 (Locking
rod correction instrument, length
300 mm/410 mm) in the sleeve of
CS 3361-01/-05 (Sleeve correction
instrument, length 300 mm/410 mm)
and push it all the way forward.
The hook of CS 3361-02/-06
(Locking rod correction instrument,
length 300 mm/410 mm)
looks ahead of the sleeve
of CS 3361-01/-05 (Sleeve
correction instrument, length
300 mm/410 mm).