Chapter 4. Specifications
4.5 Pipe Size & Materials

4.5.1 Wetted Transducers

4.5.1a
Materials
All metals and most plastics. (Consult the factory for concrete, glass and cement.)
4.5.1b
Pipe Sizes
Transit-Time Mode:
(ID) 1 to 200 in. (25.4 mm to 5 m) and larger.
4.5.2 Clamp-On Transducers
4.5.2a
Materials
All metals and most plastics. (Consult the factory for concrete, composite materials and highly corroded or lined pipes.)
4.5.2b
Pipe Sizes
Transit-Time Mode:
Outside diameter (OD) 0.5 to 300 in. (12.7 mm to 7.6 m) and larger.
TransFlection Mode:
Outside diameter (OD) 2 to 200 in. (5 cm to 5 m) and larger.
Pipe Wall Thickness
Up to 3 in. (76.2 mm)

4.6 PC Interface Software

PanaView™ software option links the DF868 to a PC. Software package includes CD and manual.