10. MAINTENANCE & FAULT-FINDING

10.1 . Cleaning

When installing the unit, handle with lint free gloves and degrease the unit to reduce the
build up of deposits.
If there is any build up of deposit on the unit, it should be gently cleaned with a cloth,
moistened with soft detergent. Solvents should not be used, and care should be taken to
avoid scratching any surfaces. The unit must be allowed to defrost naturally after being
exposed to snow or icy conditions, do NOT attempt to remove ice or snow with a tool.
Do NOT remove black "rubber" transducer caps.

10.2. Servicing

There are no moving parts or user-serviceable parts requiring routine maintenance.
Opening the unit or breaking the security seal will void the warranty and the calibration.
In the event of failure, prior to returning the unit to your authorised Gill distributor, it is
recommended that:
 All cables and connectors are checked for continuity, bad contacts, corrosion etc.
 A bench test is carried out as described in Section 10.6.
 You contact your supplier for advice

10.3. Fault-finding

Symptom
Solution
Check DC power to WindObserver II (Pipe Mount), cable and
connections.
Check comms settings of WindObserver II (Pipe Mount) (as
detailed in Section 9) and host system match, including correct
Com port
No output
Check unit is in Continuous mode
Check that in-line communication devices are wired correctly.
NOTE: It is usual for Anemometer TX + to be connected to
converter device RX +
Check comms settings of WindObserver II (Pipe Mount) and
host system match.
Corrupted output
Try a slower baud rate.
Check cable lengths and type of cable.
One way
Check wiring is in accordance with the manual.
communication
Failed / Incorrect
Check that transducer path is not blocked
WindObserver II
(Pipe Mount)
output, data
invalid flag
________________________________________________________________________________________________
WindObserver II Pipe Mount
Doc. No. 1390-PS-0009