DigiBlock – User Manual
before connecting the power supply cord.
Step 3: after the power supply cord is properly connected, switch ON the unit.
Optimized design has significantly improved the performance depending on installation
conditions of the unit. Adequately selected installation environment will certainly be helpful to
ensure longer service life and more efficient performance.
Avoid high temperature and high humidity
Never place the unit in a location where excessive heat, moisture or corrosive materials are
present. The unit should be located in a clean environment where ambient temperature is
between 15°C and 35°C and the humidity is between 45% and 80% respectively.
Avoid interference of strong electro-magnetic fields
Please keep the unit away from installations which release strong electro-magnetic field or
high-frequency waves.
Keep it away from corrosive gas
The unit should not be placed in a place with excessive corrosive gases such as chlorine,
hydrochloric acid, hydrogen sulfide, sulfurous acid gas, etc.
Place the unit on a stable platform
The installation platform must be horizontal and stable without vibration, capable of supporting
the total weight of the main unit and accessories. A distance at least 15 cm must be kept
between the left side of the instrument and the wall for undisturbed ventilation. Enough space
around the unit fan should be left for smooth ventilation.
Avoid sharing power socket with other electric appliances
An independent power socket should be prepared for the unit. The power socket should not be
shared with other electric appliances and the power supply should be grounded.
Do not place the unit in a place exposed to direct sunlight.
Do not place the unit in a dusty environment.
Attention Please: improper working environment may significantly affect the measurement
performance and reduce the life-span of the unit.
POWER SUPPLY
DigiBlock is available for power supply of 110V or 230V. The power supply of the unit should
correspond with the factory-set value before switching on, otherwise the unit will be damaged.
If the operator wants to change this value, please contact the Labtech Service Team.