b) Enter in to Method Edit interface, touch Soil and Method1
c)
Step 01 : Touch the Temp. Text box, input 100 via the digital key. Then touch Time
text box to input -2 which means heat in max. heating speed (Normally, the
average heating speed is around 5℃/min).
Step 02 : Touch the Temp. text box and input 100 via the digital key. Then touch
Time text box to input 0 which means hold the temperature at previous step
setpoint.
Touch SAVE to save the method.
d) Touch Method Check to review the method.
DigiBlock – User Manual
15
Need help?
Do you have a question about the DigiBlock ED-S Series and is the answer not in the manual?
Questions and answers