Labtech DigiBlock ED-S Series User Manual page 11

Sample preparation system
Table of Contents

Advertisement

Press
and
The unlock procedure is finished.
Attention Please: It is prohibited to change any other parameters that are not indicated in user manual.
iTouch Separate CONTROLLER
LabTech iTouch controller operates by simply touching the keys on the screen.
Connect the iTouch controller to the back of DigiBlock main unit, switch ON the DigiBlock and
the iTouch screen is automatically ready for operations.
1. Temperature control range: room temperature to 210°/415°C (based on different DigiBlock
model).
2. Working time: from 1 to 15 heating steps. Each step can be set from -2 to 9999min.
"-2" Maximum heating speed to reach temperature setpoint
"-1" End the program
"0" Maintain previous step time setting
3. Temperature sensor mode: built-in K type sensor or optional PT100 external sensor.