Labtech DigiBlock ED-S Series User Manual page 16

Sample preparation system
Table of Contents

Advertisement

e) Touch Soil then Method1 to check if the program heating curve you just set is
correct. Touch Save to quit the interface and select this program.
f)
In home page, touch RUN to start the selected program.
Edit another soil digestion method in Soil category, heat from room
temperature to 100 ℃ within 15min, hold 100 ℃ for 60min, then increase the
temperature to 150 ℃ within 8min, hold 150 ℃ for 60min. End the program.
a) In home page, touch Setup to enter into Method Edit interface.
b) Touch Soil and Method2
Step 01 : Touch the Temp. Text box, input 100 via the digital key. Then touch
Time text box to input 15.
NOTE : Recommended temperature increasing Time is '-2' (the average heating speed is around
5 ℃ /min), thus the unit can heat in the max. heating speed.
Step 02 : Touch the Temp. text box and input 100 via the digital key. Then touch
Time text box to input 60 which means hold the temperature at previous step
setpoint for 60min.
Step 03: Touch the Temp. Text box, input 150 via the digital key. Then touch Time
text box to input 8 which means temperature increase from 100C to 150C within
DigiBlock – User Manual
16

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for Labtech DigiBlock ED-S Series