Configuring the 4-20 mA
Analog Input
The Test Submenu
The Trim Submenu
Advanced Programming
To configure the 4-20 mA analog input, access the setup menu as
described on page 5-4 and complete the following steps:
Range
Cal
Setup
Serial Number
4-20mA-Input
Factory Calibration
To test the analog input, program a test value into the XDP analog
output (see page 5-6) and connect the analog output signal to the
analog input terminals. Then, proceed as follows:
4-20mA-Input
Test
Trim
At the next prompt, the mA value of the test signal is shown.
4-20mA-Input Test
mA Input Test
x.xxx
OK
To trim the analog input, make sure the XDP analog output has been
calibrated (see page 5-5) and connect the XDP analog output signal to
the analog input terminals.
4-20mA-Input
Test
Trim
The XDP analog output is now driven to exactly 4 mA.
4-20mA-In..4mA-Trim
xx.xxx
s
xx
t
January 2008
Use the [ ] and [ ] keys to
[4-20mA-Input]
select
and press
[ENTER]
.
Use the [ ] and [ ] keys to
[Test]
[ENTER]
select
and press
After noting the mA value of
the analog input signal, press
[ENTER]
to continue.
Use the [ ] and [ ] keys to
[Trim]
[ENTER]
select
and press
Press the [ ] and [ ] keys
until the meter reads exactly 4
mA. When done, press
[ENTER]
.
.
.
5-15