Purpose
Type
Command Syntax
Command Parameters
*RST Value
Query Syntax
Query Parameters
Query Response
Description
Examples
Related Commands
VM4016 Command Dictionary

OUTPut:POLarity:EXTernal:LATChed

Sets the latched interrupt output polarity on the front panel
Setting
OUTPut:POLarity:EXTernal:LATChed <polarity>
<polarity> = NORMal | INVert
NORMal
OUTPut:POLarity:EXTernal:LATChed?
N/A
ASCII string = NORM ⏐ INV
The OUTput:POLarity:EXTernal:LATChed command sets the polarity of the front
panel latched interrupt output. When the polarity is set for normal, the output will be
high when there is an interrupt event. When set for invert, the output will be low when
there is an interrupt event.
Command / Query
OUTP:POL:EXT:LATC INV
OUTP:POL:EXT:LATC?
None
Response (Description)
(Sets the polarity of the front panel latched interrupt
output to inverted)
INV (Returns the value for the front panel latched
interrupt output)