Purpose
Type
Command Syntax
Command Parameters
*RST Value
Query Syntax
Query Parameters
Query Response
Description
Examples
Related Commands
72
INPut:DEBounce
Sets the debounce time
Setting
INPut:DEBounce <value>
<value> = 9.6 μs to 0.6291456 s
19.2 μs
INPut:DEBounce?
N/A
Numeric ASCII value from 0.0000096 to 0.6291456
The INPut:DEBounce command sets the time period for the digital debounce circuitry.
By programming a debounce time of 1 ms, an input must exceed its threshold level for
a period of 1 ms before it is recognized as a valid input. The debounce resolution is
9.6 μs. The debounce time set is applied to all channels.
Command / Query
INP:DEB 9.6e-6
INP:DEB
None
Response (Description)
(Sets a digital debounce time of 9.6 μs)
0.0000096 (Indicates that the debounce time is set to
9.6 μs)
VM4016 Command Dictionary
VXI Technology, Inc.