INPUT:RANGE

The input range command selects the input range of one or more channels. The input range may
be either set for ±10 V or ±100 V. It is important to note that the input offset is normalized to
±10 V range. The actual input offset in the 100 V range is ten times the set value.
INPut:RANGe <range>,<channel_list>
EXAMPLES
INPut:RANGe 100,(@1:16)
INPut:RANGe? 7
100
INP:RANG 10,(@4:6)
INP:RANG? 5
10
VM4016 Programming
Where <range> is 10 V | 100 V.
Where <channel_list> is the standard
channel list format supporting channels 1
through 16.
Sets the input range for Channels 1 through
16 to 100 V.
Returns the input range for Channel 7 as
100 V.
Sets the input range for Channels 4 through
6 to 10 V.
Returns the input range for Channel 5 as
10 V.