Keysight 16200B External DC Bias Adapter Operation and Service Manual...
Page 2
A CAUTION notice denotes a haz- Published by: eral Acquisition Regulation (“FAR”) ard. It calls attention to an oper- 2.101. Pursuant to FAR 12.212 and Keysight Technologies International 27.405-3 and Department of Defense ating procedure, practice, or the Japan G.K, FAR Supplement (“DFARS”)
1 General Information Introduction The purpose of this manual is to enable you to use your 16200B External DC Bias Adapter efficiently and confidently. This manual contains both general and specific information. To use the 16200B to perform a specific function (without having to read the entire manual), follow the directions in "Using the...
General Information Accessories Supplied Accessories Supplied The following accessories are supplied with the 16200B: Table 1-1 Furnished Accessories Description Part Number Quantity Operation and Service Manual Option ABA Options Available The Option 001 is supplied for the 16200B: Table 1-2...
Electrostatic discharge (ESD) can damage the highly sensitive microcircuits in the 16200B External DC Bias Adapter. ESD damage is most likely to occur as the 16200B is being connected or disconnected. Protect the 16200B from ESD damage by wearing a grounding strap that provides a low resistance path to ground.
General Information Specifications Specifications This section lists the complete 16200B specifications. These specifications are the performance standards and limits against which the 16200B is tested. When shipped from the factory, the 16200B meets the following specifications: External DC Bias Up to 40 V can be applied to the DC ±...
Preparation for Use Introduction 2 Preparation for Use Introduction This chapter explains how to install the 16200B External DC Bias Adapter. The topics covered include initial inspection, ambient environmental considerations, connecting the adapter for use, and repacking the adapter. Initial Inspection The adapter has been carefully inspected electrically and mechanically before being shipped from the factory.
Page 10
LOAD Device (0.6 mm × 0.3 mm) 0699-6926 LOAD Device (1 mm × 0.5 mm) 5182-0433 LOAD Device (1.6 mm × 0.8 mm) 5182-0434 LOAD Device (2.0 mm × 1.25 mm) 5182-0435 LOAD Device (3.2 mm × 1.6 mm) 5182-0436 Keysight 16200B External DC Bias Adapter...
Environmental Considerations Operating and Storage The 16200B should be operated within an ambient temperature range of 0 %C to 55 %C and relative humidity up to 95% RH at 40 %C (non-condensing). The 16200B may be stored within a temperature range of 40 %C to +70 %C, and −...
3. DC Source Input. This terminal is connected to an external current source in order to apply dc bias current to the DUT. 4. V Monitor. This terminal is connected to a Volt Meter to monitor the dc bias level. Keysight 16200B External DC Bias Adapter...
Alternatively, comparable packaging materials may be used. Wrap the adapter in heavy paper and pack in anti-static plastic packing material. Use sufficient shock absorbing material on all sides of the 16200B to provide a thick, firm cushion and to prevent movement. Seal the shipping container securely and mark it FRAGILE.
Page 14
Preparation for Use Repacking the Adapter Keysight 16200B External DC Bias Adapter...
After the calibration of the test station has been completed, connect the connector on the underside of the 16200B to the 7 mm connector on the test station and connect the connector on the underside of the test fixture to the 7 mm connector on the upper side of the 16200B.
Operation Using the 16200B with the 4291B Compensation Procedure LOAD Fixture Compensation LOAD Compensation corrects the phase-shift induced error, making the Electrical Length Compensation unnecessary. From the LOAD devices supplied with the adapter, select a LOAD with an impedance value that is accurately known and stable and that is nearest in size to the DUT.
1. Option E4990A-120 is required. 2. Option 005 is required. Calibration Calibration must be performed on the 16200B DC Bias Adapter before the Test Fixture is attached. Execute the calibration in accordance with the Operation Manual for the measuring instrument you are using.
Also, because of its location, this capacitor makes it impossible to bias samples from the measuring instrument's internal bias source. Thus, the 16200B can not be used for applications in which the instrument's internal bias source is used.
Where the dc bias current applied: I - - - - - - - Where f is measurement frequency in MHz, Zx is impedance of DUT in I is current in A. Keysight 16200B External DC Bias Adapter...
Page 20
Operation DC Bias Keysight 16200B External DC Bias Adapter...
RoHS support part backward incompatible with the non-RoHS 16200B. Special handling is needed while using the RoHS replacement part on the non-RoHS 16200B. The original support part number is replaced by the respective “RoHS Compliant Keysight Part No.”.
Page 22
Service Replaceable Parts Figure 4-1 Replaceable Parts Identifiers Keysight 16200B External DC Bias Adapter...
Page 23
DC Bias level) 2110-0565 2110-0565 Fuse Cover 2110-0056 2110-0056 Fuse (6A) 1250-0907 1250-0907 CONTACT CENTER Schematic Diagram Figure 4-2 shows the schematic diagram of the 16200B. Figure 4-2 16200B Schematic Diagram Keysight 16200B External DC Bias Adapter...
Page 24
Service Schematic Diagram Keysight 16200B External DC Bias Adapter...