27.401 or DFAR 227.7103-5 (c), as applicable adhered to, could result in personal injury or in any technical data. death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met. Keysight 11904S User’s and Service Guide...
Certification Keysight Technologies certifies that this product met its published specifications at the time of shipment from the factory. Keysight Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST, formerly NBS), to the extend allowed by the Institute’s calibration facility, and to the calibration facilities of the other International Standards Organization members.
To contact Keysight for sales and technical support, refer to the support links on the following Keysight websites: – www.keysight.com/find/adapters (product-specific information and support, software and documentation updates) – www.keysight.com/find/assist (worldwide contact information for repair and service) Keysight 11904S User’s and Service Guide...
....... . . 36 Keysight 11904S User’s and Service Guide...
Page 10
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight 11904S User’s and Service Guide...
Page 11
Keysight 11904S 2.4 mm/2.92 mm Adapter Set User’s and Service Guide General Information Adapter Set Overview Incoming Inspection Clarifying the Sex of Connectors Preventive Maintenance This chapter provides you with an overview of the 11904S adapter set.
General Information Adapter Set Overview The Keysight 11904S adapter set is designed to give network analyzer systems with 2.4 mm test ports (such as the Keysight 8510, 8722, and the PNA Series) the ability to perform measurements on devices with 2.92 mm connectors. The set can be used to achieve calibrated measurements of 2.92 mm devices up to 40 GHz, and 2.4 mm devices up to 50 GHz.
Keysight will arrange for repair or replacement of incomplete or damaged shipments without waiting for a settlement from the transportation company. See “Returning a Kit or Device to Keysight” on page 49. Keysight 11904S User’s and Service Guide...
A connector gage is referred to in terms of the connector that it measures. For instance, a male connector gage has a female connector on the gage so that it can measure male devices. Keysight 11904S User’s and Service Guide...
Improper connections, resulting from pin depth values being out of the observed limits (see Table 2-2 on page 20) or from bad connection techniques, can also damage these devices. Keysight 11904S User’s and Service Guide...
Page 16
General Information THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight 11904S User’s and Service Guide...
Page 17
Keysight 11904S 2.4 mm/2.92 mm Adapter Set User’s and Service Guide Specifications Environmental Requirements Mechanical Characteristics Electrical Specifications This chapter provides the environmental requirements, mechanical characteristics, and electrical specifications of the 11904S adapter set.
Specifications Environmental Requirements The Keysight 8497K Attenuator is designed to fully comply with Keysight Technologies’ product operating environmental specifications as shown in Table 2-1. Table 2-1 Environmental specifications Parameter Limits Temperature – Operating – +20 °C to +26 °C – Storage –...
Mechanical characteristics such as center conductor protrusion and pin depth are not performance specifications. They are, however, important supplemental characteristics related to electrical performance. Keysight Technologies verifies the mechanical characteristics of the devices in the kit with special gaging processes and electrical testing.
Page 20
When measuring pin depth, the measured value (resultant average of three or more measurements) is not the true value. Always compare the measured value with the observed pin depth limits in Table 2-2 to evaluate the condition of device connectors. Keysight 11904S User’s and Service Guide...
Certification Keysight Technologies certifies that this product met its published specifications at the time of shipment from the factory. Keysight further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST) to the extent allowed by the institute’s calibration facility, and to the calibration facilities of other International...
Page 22
Specifications THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight 11904S User’s and Service Guide...
Page 23
Use, Maintenance, and Care of the Devices Electrostatic Discharge Visual Inspection Calibration Information Cleaning the Connectors Gaging Connectors Gaging Procedures Connections Operational Check Handling and Storage This chapter teaches you how to use, maintain, and provide care for the 11904S adapter set.
2 While wearing a grounded wrist strap, grasp the outer shell of the cable connector. 3 Connect the other end of the cable to the test port. 4 Remove the short from the cable. Figure 3-1 ESD protection setup Keysight 11904S User’s and Service Guide...
If a connector shows deep scratches or dents, particles clinging to the mating plane surfaces, or uneven wear, clean and inspect it again. Devices with damaged connectors should be discarded. Determine the cause of damage before connecting a new, undamaged connector in the same configuration. Keysight 11904S User’s and Service Guide...
The precision slotless connector was developed to eliminate this problem. The PSC has a center conductor with a solid cylindrical shell, the outside diameter of which does not change when mated. Instead, the center conductor has an internal contact that flexes to accept the male pin. Keysight 11904S User’s and Service Guide...
Use, Maintenance, and Care of the Devices Calibration Information The calibration procedure for using the Keysight 11904S adapters requires that a 2.4 mm calibration be done using the calibration definitions of the 11904S instead of the calibration definitions of the Keysight 85056A 2.4 mm calibration kit. Although response, 1-Port, and 1-Path 2-Port calibrations may also be done, the following section is a general overview of a full 2-Port calibration, as it is the most complex of the calibrations.
VNA has an earlier version of the calibration kit definition file, obtain a copy of the latest file from Keysight Technologies. Figure 3-2 Full two-port calibration - Keysight 8510 VNA using calibration kit definition CK_292MMA3 and higher Keysight 11904S User’s and Service Guide...
Page 29
7 If the 2.92 mm device being tested is a non-insertable, swap out one of the 2.4 mm to 2.92 mm adapters with one that will mate with the test device. Keysight 11904S User’s and Service Guide...
Clean the connector threads. c Let the alcohol evaporate, then blow the threads dry with a gentle stream of clean, low- pressure compressed air or nitrogen. Always completely dry a connector before you reassemble or use it. Keysight 11904S User’s and Service Guide...
Page 31
Always completely dry a connector before you reassemble or use it. 4 Inspect Inspect the connector to make sure that no particles or residue remain. Refer to “Visual Inspection” on page 25. Keysight 11904S User’s and Service Guide...
Use, Maintenance, and Care of the Devices Gaging Connectors The gages available from Keysight Technologies are intended for preventive maintenance and troubleshooting purposes only. They are effective in detecting excessive center conductor protrusion or recession and conductor damage on DUTs, test accessories, and the calibration kit devices.
Refer to “Final connection using a torque wrench” on page 36 for additional information. c Gently tap the barrel of the gage with your finger to settle the gage reading. Keysight 11904S User’s and Service Guide...
Page 34
Compare the average reading with the observed pin depth limits in Table 2-2 on page 20. Keysight 11904S User’s and Service Guide...
90 N-cm (8 in-lb) ±9.0 N-cm (±0.8 in-lb) Using a torque wrench guarantees that the connection is not too tight, preventing possible connector damage. It also guarantees that all connections are equally tight each time. Keysight 11904S User’s and Service Guide...
Page 37
You don’t have to fully break the handle of the torque wrench to reach the specified torque; doing CAUTION so can cause the handle to kick back and loosen the connection. Any give at all in the handle is sufficient torque. Keysight 11904S User’s and Service Guide...
2 Use another open-end wrench to loosen the connecting nut. 3 Complete the separation by hand, turning only the connecting nut. 4 Pull the connectors straight apart without twisting, rocking, or bending either of the connectors. Keysight 11904S User’s and Service Guide...
– 16 averages – 2.4 mm sliding load cal Figure 3-6 2.4 mm to 2.92 mm adapter to lowband load 2 Connect the 2.4 mm -f- to 2.92 mm -f- adapter to the calibrated test port. Keysight 11904S User’s and Service Guide...
Page 40
5 Remove the lowband load and connect a 2.4 mm sliding load -f- to the adapter. Refer to the 85056A 2.4 mm Precision Calibration Kit for information on connecting a sliding load. The return loss should be > 20 dB from 4GHz to 40 GHz. Keysight 11904S User’s and Service Guide...
– Do not set connectors contact-end down on a hard surface. The plating and the mating plane surfaces can be damaged if the interface comes in contact with any hard surface. Keysight 11904S User’s and Service Guide...
Keysight 11904S 2.4 mm/2.92 mm Adapter Set User’s and Service Guide Performance Verification Introduction How Keysight Verifies the Devices in Your Kit Recertification This chapter provides information on the recertification steps for the 11904S adapter set.
The performance of your adapter set can only be verified by returning the kit to Keysight Technologies for recertification. The equipment and calibration standards required to verify the specifications of the devices in the kit have been specially manufactured and are not commercially available. Keysight 11904S User’s and Service Guide...
These two steps establish a traceable link to NIST for Keysight to the extent allowed by the institute’s calibration facility. The specifications data provided for the devices in the kit is traceable to NIST through Keysight Technologies. Keysight 11904S User’s and Service Guide...
NOTE date. Where to send a kit for recertification Contact Keysight Technologies for information on where to send your kit for recertification. Refer to “Returning a Kit or Device to Keysight” on page 49 for details on sending your kit.
Page 47
Keysight 11904S 2.4 mm/2.92 mm Adapter Set User’s and Service Guide Troubleshooting Troubleshooting Process Returning a Kit or Device to Keysight How Keysight Verifies the Devices in Your Kit Recertification This chapter provides basic troubleshooting instructions for the 11904S adapter set.
This manual contains limited information about network analyzer system operation. For complete information, refer to the instrument documentation. If you suspect a bad calibration, or if your network analyzer does not pass performance verification, follow the steps in Figure 5-1. Figure 5-1 Troubleshooting flowchart Keysight 11904S User’s and Service Guide...
Returning a Kit or Device to Keysight If your kit or device requires service, contact Keysight Technologies for information on where to send it. See the end of this manual for contact information. Include a service tag (located near the end of this manual) on which you provide the following information: –...
These two steps establish a traceable link to NIST for Keysight to the extent allowed by the institute’s calibration facility. The specifications data provided for the devices in the kit is traceable to NIST through Keysight Technologies. Keysight 11904S User’s and Service Guide...
NOTE date. Where to send a kit for recertification Contact Keysight Technologies for information on where to send your kit for recertification. Refer to “Returning a Kit or Device to Keysight” on page 49 for details on sending your kit.
Page 52
Troubleshooting THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight 11904S User’s and Service Guide...