Keysight Technologies 85130C Instruction Manual
Keysight Technologies 85130C Instruction Manual

Keysight Technologies 85130C Instruction Manual

Nmd 3.5 mm to type-n adapter kit

Advertisement

Keysight 85130C
NMD 3.5 mm to
Type-N Adapter Kit
Operation and
Service Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 85130C and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Keysight Technologies 85130C

  • Page 1 Operation and Service Manual Keysight 85130C NMD 3.5 mm to Type-N Adapter Kit...
  • Page 2: Safety Notices

    DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY beyond those set forth in the TERMS IN THE SEPARATE EULA shall apply, except to the © Keysight Technologies, Inc. AGREEMENT WILL CONTROL. extent that those terms, rights, or 1987-2017 licenses are explicitly required...
  • Page 3: General Information

    Description The 85130C adapters are used when type-N devices must be attached to a 3.5 mm test set in a network analyzer system. The test set end of the adapters has a NMD-3.5 mm (f) connector while the Device Under Test (DUT) end has a precision slotless type-N connector (PSC-N).
  • Page 4 85130C Adapter Kit The allowable pin position for type-N connectors is: male = –0.2075 to –0.2079 inch (–5.2705 to –5.28066 mm) female = +0.2061 to +0.2065 inch (+5.23494 to +5.2451 mm) In a mated pair of type-N connectors, the center conductor’s mating plane is offset from the outer conductor’s mating plane by 0.207 inch (5.2578 mm) in the direction of the male connector.
  • Page 5: Proper Use

    85130C Adapter Kit Proper Use Attach the adapters to the test ports and tighten them finger tight. Use the spanner wrench to hold the test set end of the adapter and torque the test set connector with a 20 mm torque wrench set to 96 N-cm (8 in-lb).
  • Page 6 85130C Adapter Kit 4. Set an IF bandwidth of 100 Hz. 5. Perform, and then save, a 3.5 mm 1-port S11 calibration. 6. With correction turned on, select the time domain mode. 7. Set a stimulus start time for the sweep to –0.05 nano-seconds.
  • Page 7: Replaceable Parts

    Equipment and Supplies The following equipment and supplies are required for the maintenance and use of, but are not supplied with, your 85130C adapter. kit. Item Part Number 3.5 mm gage kit (part of the 85052B calibration kits)
  • Page 8 85130C Adapter Kit Operating and Service Manual 85130-90010...
  • Page 9 Installation Note Xxxxx-xxxxx...
  • Page 10 This information is subject to change without notice. © Keysight Technologies 1987-2017 Edition 2, April 2017 www.keysight.com...