A.03.05 or later be installed on the Infiniium DCA. The Infiniimax Active Probes include the 1131A, 1132A, and 1134A. Other supported active probes include the 1152A, 1156A, 1157A, and 1158A. When using the N1022B, always work at a static-safe work station and wear a wrist-strap CAUTION or heel-strap.
No operator serviceable parts inside. Refer servicing to qualified personnel. To prevent electrical WARNING shock, do not remove covers. To prevent electrical shock, d isconnect the N1022B from probe power before cleaning. Use a dry cloth WARNING to clean the external case parts. Do not attempt to clean internally.
N1022B Probe Adapter Installation Category Installation Categories (overvoltage categories) are determined by the transient overvoltage levels that may be expected. CAT I: Mains isolated. CAT II: Line voltage in appliance and to wall outlet. CAT III: Line voltage behind wall outlet to next level of distribution.
N1022B Probe Adapter Using the Probe Adapter There are a few things to consider when connecting the N1022B Active Probe Adapter to an active probe and the Infiniium DCA. - The dynamic range of the system will be 3.2V (6.4 Vp-p), which with probe offset, covers most digital technologies.
Be sure that the instrument is properly earth-grounded to prevent buildup of static charge. The N1022B can be directly connected to any 86100-series or 83480-series module with a 3.5 mm NOTE electrical channel input.
Page 10
However, the technologies necessary to achieve high performance do not allow the probe to be unbreakable. Treat the probe adapter with care. During operation, the N1022B probe adapter’s temperature may feel warm when you touch it. This is NOTE normal and does not indicate a problem.
Page 11
N1022B Probe Adapter Figure 1 Probe Channel Dialog Box Perform the probe calibration procedure so that the DC gain, offset zero, and offset gain will be calibrated. You can access information about probe attenuation and probe calibration in the Infiniium DCA Help system.
N1022B Probe Adapter Static-Safe Workstation Electrostatic discharge (ESD) can damage or destroy electronic components. All work on electronic assemblies should be performed at a static-safe workstation. The following figure shows an example of a static-safe work station using two types of ESD protection: - Conductive table-mat and wrist-strap combination.
Page 13
N1022B Probe Adapter These techniques for a static-safe work station should not be used when working on circuitry with a WARNING vol tage potential greater than 500 vol ts. The following suggestions may help reduce ESD damage that occurs during testing and servicing operations.
N1022B Probe Adapter Specifications and Regulatory Information This section lists specifications and characteristics and regulatory information of the probe adapter. Specifications The distinction between specifications and characteristics is described as follows: ° - Specifications describe warranted performance over the temperature range 0 C to °...
Page 15
Weight 200 g (7 oz) max Dimensions (excludes power cable) Height 34 mm (1.34 in.) Weight 38 mm (1.50 in.) Length 61 mm (2.4 in.) Power Cable Length 200 mm (7.9 in.) from connector tip to N1022B body User’s Guide...
Page 16
N1022B Probe Adapter Table 2 Performance Specifications Item Specification Frequency Range (Characteristic) 0 - 18 GHz Power Input +17V 225 mA –17V 225 mA Output Voltage +12V ±5% –12V ±5% V prog ±5% –V prog ±5% Output Current 0 - 200 mA (+12V) 0 - 200 mA (–12V)
N1022B Probe Adapter Figure 3 N1022B I/O Connections Regulatory Information Electrostatic Discharge Immunity - IEC 61000-4-2: 1995 passes criterion B. This product complies with the Electrostatic Discharge immunity requirement in IEC/EN 61326 using performance criterion B. Degradation of some product specifications can occur during the instance of an electrostatic discharge.
Page 18
N1022B Probe Adapter Table 3 Notice for Germany: Noise Declaration Acoustic Noise Emission Geraeuschemission LpA < 70 dB LpA < 70 dB Operator position am Arbeitsplatz Normal position normaler Betrieb per ISO 7779 nach DIN 45635 t.19 User’s Guide...
N1022B Probe Adapter Service Service Keysightor the N1022B Active Probe Adapter is by replacement only. To order a new N1022B, contact your local Keysight Technologies Sales Office. Return the failed probe adapter to your local Keysight service center. If the probe adapter is still under warranty or is covered by an Keysight maintenance contract, it will be replaced under the terms of the warranty or contract.
Page 20
N1022B Probe Adapter Table 4 Probe Power Outputs Supply Power Output –3V Offset 5 and ring Probe ID –12V +12V User’s Guide...
N1022B Probe Adapter Returning the N1022B When the probe adapter is returned to an Keysight service office for servicing, it must be adequately packaged and have a complete description of the failure symptoms attached. Please notify the service office before returning your probe adapter for service. Any special arrangements for the adapter can be discussed at this time.