Keysight Technologies 87300B Startup Manual
Keysight Technologies 87300B Startup Manual

Keysight Technologies 87300B Startup Manual

Directional couplers, hybrid coupler

Advertisement

Quick Links

Keysight 87300B/C/D,
87301B/C/D/E Directional
Couplers, 87310B Hybrid
Coupler
Startup Guide

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 87300B and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Keysight Technologies 87300B

  • Page 1 Keysight 87300B/C/D, 87301B/C/D/E Directional Couplers, 87310B Hybrid Coupler Startup Guide...
  • Page 2 DOCUMENT IS PROVIDED “AS IS,” sition Regulation (“FAR”) 2.101. Pursu- AND IS SUBJECT TO BEING © Keysight Technologies 2013 - 2019 ant to FAR 12.212 and 27.405-3 and CHANGED, WITHOUT NOTICE, IN No part of this manual may be repro- Department of Defense FAR Supple- FUTURE EDITIONS.
  • Page 3: Waste Electrical And Electronic Equipment (Weee) Directive

    Waste Electrical and Electronic Equipment (WEEE) Directive This instrument complies with the WEEE Directive marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste. Product category: With reference to the equipment types in the WEEE directive Annex 1, this instrument is classified as a “Monitoring and Control Instrument”...
  • Page 4 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight 87300 Series Startup Guide...
  • Page 5: Table Of Contents

    ..........9 Keysight 87300B/C/D and 87301B/C/D/E Directional Coupler .
  • Page 6 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight 87300 Series Startup Guide...
  • Page 7: List Of Figures

    List of Figures Figure 1 Equipment setup using network analyzer with directional coupler ........10 Figure 2 Equipment setup using network analyzer with directional coupler port 1 and port 3...
  • Page 8 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight 87300 Series Startup Guide...
  • Page 9: Introduction

    Introduction This startup guide describes the procedure for making quick check on 87300B, C, D, 87301B, C, D, E directional couplers and 87310B hybrid coupler.Throughout this quick check procedure, each connection made must be accomplished by following the Principles of Microwave Connector Care. Select the correct microwave cables and adapters when preparing the test setup.
  • Page 10: Keysight 87300B/C/D And 87301B/C/D/E Directional Coupler

    Keysight 87300B/C/D and 87301B/C/D/E Directional Coupler 1 Calibrate the network analyzer with full 2-Port calibration with setup based on network analyzer model being used.The test setup is shown in Figure Figure 1 Equipment setup using network analyzer with directional coupler 2 Save Cal Set to network analyzer’s memory.
  • Page 11: Equipment Setup Using Network Analyzer With Directional Coupler Port 1 And Port

    Figure 2 Equipment setup using network analyzer with directional coupler port 1 and port 3 6 Measure nominal coupling (S31/S21) and frequency sensitivity (S31/ S21) by using the equations on page 7 With the same setup as Figure 2, normalize S31 8 Reverse the coupler as shown in Figure 3.
  • Page 12: Equipment Setup Using Network Analyzer With Directional Coupler

    Figure 3 Equipment setup using network analyzer with directional coupler port 2 and port 3 min. coupling point + max. coupling point Nominal coupling = min. coupling point - max. coupling point Frequency sensitivity = From the current measurement trace (after Data / Mem function performed), turn on Marker 1 and 2, then locate the minimum and maximum coupling points respectively.
  • Page 13: Keysight 87310B Hybrid Coupler

    Keysight 87310B Hybrid Coupler 1 Select the correct microwave cables and adapters when preparing the test setup. 2 Calibrate the network analyzer with full 2-port calibration with setup based on network analyzer model being used. The test setup is shown in Figure Figure 4 Equipment setup using network analyzer with hybrid coupler...
  • Page 14 Figure 5 Equipment setup using network analyzer with hybrid coupler port 1 and port 3 7 Measure amplitude imbalance (S31/ S21) The following example (Figure 6), of the maximum deviation from 0.0 dB reference is approximately -0.7 dB at 1 GHz. The allowable maximum imbalance is 1.0 dB So the measured amplitude imbalance value -0.7 dB is within the specification.
  • Page 15 Figure 6 Amplitude imbalance 8 With the same setup as above, change the format to phase and measure the phase imbalance (S31/S21) 9 Connect the DUT as shown in Figure 7. Measure and record isolation S41 and S44 SWR. [1] Phase imbalance is defined as the difference between phases of port 3 to port 2 within the frequency band.
  • Page 16: Equipment Setup Using Network Analyzer With Hybrid

    Figure 7 Equipment setup using network analyzer with hybrid coupler port 1 and port 4 10 Connect the DUT as shown in Figure 8. Measure and record isolation S32. Figure 8 Equipment setup using network analyzer with hybrid coupler port 1 and port 3 Keysight 87300 Series Startup Guide...
  • Page 17 This information is subject to change without notice. Always refer to the English version at the Keysight website for the latest revision. © Keysight Technologies 2013 - 2019 Edition 1, April 3, 2019 Printed in Malaysia *87300-80014* 87300-80014...