Waste Electrical and Electronic Equipment (WEEE) Directive This instrument complies with the WEEE Directive marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste. Product category: With reference to the equipment types in the WEEE directive Annex 1, this instrument is classified as a “Monitoring and Control Instrument”...
Page 4
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight 87300 Series Startup Guide...
List of Figures Figure 1 Equipment setup using network analyzer with directional coupler ........10 Figure 2 Equipment setup using network analyzer with directional coupler port 1 and port 3...
Page 8
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight 87300 Series Startup Guide...
Introduction This startup guide describes the procedure for making quick check on 87300B, C, D, 87301B, C, D, E directional couplers and 87310B hybrid coupler.Throughout this quick check procedure, each connection made must be accomplished by following the Principles of Microwave Connector Care. Select the correct microwave cables and adapters when preparing the test setup.
Keysight 87300B/C/D and 87301B/C/D/E Directional Coupler 1 Calibrate the network analyzer with full 2-Port calibration with setup based on network analyzer model being used.The test setup is shown in Figure Figure 1 Equipment setup using network analyzer with directional coupler 2 Save Cal Set to network analyzer’s memory.
Figure 2 Equipment setup using network analyzer with directional coupler port 1 and port 3 6 Measure nominal coupling (S31/S21) and frequency sensitivity (S31/ S21) by using the equations on page 7 With the same setup as Figure 2, normalize S31 8 Reverse the coupler as shown in Figure 3.
Figure 3 Equipment setup using network analyzer with directional coupler port 2 and port 3 min. coupling point + max. coupling point Nominal coupling = min. coupling point - max. coupling point Frequency sensitivity = From the current measurement trace (after Data / Mem function performed), turn on Marker 1 and 2, then locate the minimum and maximum coupling points respectively.
Keysight 87310B Hybrid Coupler 1 Select the correct microwave cables and adapters when preparing the test setup. 2 Calibrate the network analyzer with full 2-port calibration with setup based on network analyzer model being used. The test setup is shown in Figure Figure 4 Equipment setup using network analyzer with hybrid coupler...
Page 14
Figure 5 Equipment setup using network analyzer with hybrid coupler port 1 and port 3 7 Measure amplitude imbalance (S31/ S21) The following example (Figure 6), of the maximum deviation from 0.0 dB reference is approximately -0.7 dB at 1 GHz. The allowable maximum imbalance is 1.0 dB So the measured amplitude imbalance value -0.7 dB is within the specification.
Page 15
Figure 6 Amplitude imbalance 8 With the same setup as above, change the format to phase and measure the phase imbalance (S31/S21) 9 Connect the DUT as shown in Figure 7. Measure and record isolation S41 and S44 SWR. [1] Phase imbalance is defined as the difference between phases of port 3 to port 2 within the frequency band.
Figure 7 Equipment setup using network analyzer with hybrid coupler port 1 and port 4 10 Connect the DUT as shown in Figure 8. Measure and record isolation S32. Figure 8 Equipment setup using network analyzer with hybrid coupler port 1 and port 3 Keysight 87300 Series Startup Guide...