Keysight Technologies 16065C Operation And Service Manual
Keysight Technologies 16065C Operation And Service Manual

Keysight Technologies 16065C Operation And Service Manual

External bias adapter

Advertisement

Quick Links

Keysight 16065C External Bias Adapter
Operation and
Service Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 16065C and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Keysight Technologies 16065C

  • Page 1 Keysight 16065C External Bias Adapter Operation and Service Manual...
  • Page 2 A CAUTION notice denotes a haz- Published by: eral Acquisition Regulation (“FAR”) ard. It calls attention to an oper- 2.101. Pursuant to FAR 12.212 and Keysight Technologies International 27.405-3 and Department of Defense ating procedure, practice, or the Japan G.K, FAR Supplement (“DFARS”)
  • Page 3: Table Of Contents

    Using the 16065C ........
  • Page 4 Contents Keysight 16065C External Bias Adapter...
  • Page 5: General Information

    General Information Introduction The purpose of this manual is to enable you to use your 16065C External Bias Adapter efficiently and confidently. This manual contains both general and specific information. To use the 16065C to perform a specific function (without having to read the entire manual), follow the directions in "Using the...
  • Page 6: Operating And Safety Precautions

    Electrostatic discharge (esd) can damage the highly sensitive microcircuits in your instrument. ESD damage is most likely to occur as the 16065C is being connected or disconnected. Protect the 16065C from ESD damage by wearing a grounding strap that provides a high resistance path to ground.
  • Page 7: Specifications

    Specifications This section lists the complete 16065C specifications. These specifications are the performance standards and limits against which the 16065C is tested. When shipped from the factory, the 16065C meets the following specifications: Up to ±40 V can be applied to the DC BIAS INPUT BNC External DC Bias connector.
  • Page 8 General Information Specifications Keysight 16065C External Bias Adapter...
  • Page 9: Operating And Safety Precautions

    Step 1. Inspect the shipping container for damage, and keep the shipping materials until the inspection is completed. Step 2. Verify that the shipping container contains everything shown in Figure 2-1 and listed in Table 2-1. Step 3. Inspect the exterior of the 16065C for any signs of damage.
  • Page 10: Environmental Considerations

    55 °C and relative humidity up to 95% at 40 °C (non-condensing). The 16065C may be stored within a temperature range of –40 °C to +70 °C, and at a relative humidity of up to 95% at +65 °C (non-condensing).
  • Page 11: Connecting The Adapter For Use

    Alteratively, comparable packaging materials may be used. Wrap the adapter in heavy paper and pack in anti-static plastic packing material. Use sucient shock absorbing material on all sides of the 16065C to provide a thick, firm cushion and to prevent movement. Seal the shipping container securely and mark it FRAGILE.
  • Page 12 Operating and Safety Precautions Repackaging the Adapter Keysight 16065C External Bias Adapter...
  • Page 13: Operation

    Keysight 16065C External Bias Adapter Operation and Service Manual 3 Operation Introduction This chapter describes the features of the 16065C (Figure 3-1), the connection to an external voltage source and test fixture, and compensation techniques. Figure 3-1 Adapter Features 1. Four-terminal-pair test leads. These are connected to the UNKNOWN terminals of an LCR meter or impedance analyzer.
  • Page 14: Open/Short Correction

    Step 2. Connect the test fixture directly to the 16065C. Step 3. Connect the dc voltage source to the 16065C's DC BIAS INPUT BNC. Do NOT turn on the voltage source while correction is in progress. Step 4. Perform OPEN and SHORT correction as described in the measuring instrument manual.
  • Page 15: Dc Bias

    20 mA at 40 V. 16065C must be Current Limiting The 16065C's DC Bias Input has a 2 kΩ current limiting resistor in series with the component under test. Capacitive Component Requires Time to Charge The 16065C DC BIAS INPUT has a 2 kΩ current limiting resistor in series with the component under test.
  • Page 16: Additional Error For D Measurement

    ) is derived by the following equation: _ 0.7 / f (Typical) The compensation function of the E4981A and E4980AL can correct the additional error caused by the 16065C. For more information on compensation and correction, refer your instrument's manual. Keysight 16065C External Bias Adapter...
  • Page 17: Service

    RoHS support part backward incompatible with the non-RoHS 16065C. Special handling is needed while using the RoHS replacement part on the non-RoHS 16065C. The original support part number is replaced by the respective “RoHS Compliant Keysight Part Number”.
  • Page 18 Replaceable Mechanical Parts RoHS Reference Keysight Compliant Descrption Designator Part Number Keysight Part Number 1250-0083 1250-0083 Connector BNC (for "Ext DC Bias") 2190-0016 2190-0016 Washer 2950-0043 2950-0043 16065-04002 16065-04602 Chassis 16065-04001 16065-04601 Chassis Top Cover Keysight 16065C External Bias Adapter...
  • Page 19 16065-01271 Angle 1250-0118 1250-0118 Connector BNC 16047-40002 16047-40002 Insulator 2190-0016 2190-0016 Washer 0360-1190 0360-1190 Terminal Solder Lug 2950-0043 2950-0043 0403-0427 0403-0427 Bumper Foot 1. These parts are for 1. 2. These parts are for 15. Keysight 16065C External Bias Adapter...
  • Page 20 Service Replaceable Parts Keysight 16065C External Bias Adapter...
  • Page 21 This information is subject to change without notice. © Keysight Technologies 1991 - 2019 Edition 3, August 23, 2019 *16065-90020* 16065-90020 www.keysight.com...