STATus:QUEue?
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Returns the next event code and message in the Error/Event queue.
The Error/Event queue is cleared at power on or when you send the *CLS
command. This is derived from the standard SCPI command. The SYSTem:ER-
Ror? query is an alias.
There is no command associated with this query.
STATus:QUEue[:NEXT]?
Parameter
Not Applicable
0
None
None
STAT:QUE?
Query:
Response: –221, "Settings Conflict."
STATus:QUEue:ENABle
SYSTem:ERRor?
STATus Subsystem
Query response
<NR1>, <string>