Page 3
Int rodu ct ion Dear Customer, Thank you for choosing Apera Instruments EC60-Z Smart Conductivity Tester. Please carefully read this manual before using the product in order to have a reliable testing experience. 1.1 This product is designed with a two-way control on both the tester and ZenTest Mobile App. Please refer to the functions available on each platform in the following table.
Page 4
Bat t e r y I nst a ll at i on Please install batteries according to the following steps. *Please note direction of batteries: All POSITIVE SIDES (“+”) FACING UP. (Wrong installation of batteries will cause damage to the tester and potential hazards) ①...
Co mpl et e Kit Graph-2 T hings to Know Bef o re U s e A few drops of water are added to the probe cap to keep the conductivity electrode in an activated state before the tester leaves factory. Generally, users can start using the tester directly. For conductivity electrode that hasn’t been used for a long time, users should soak the electrode in 12.88 mS calibration solution for 5-10 minutes or in tap water for 1 to 2 hours before use.
Page 6
After calibration, place the probe in 12.88 mS/cm conductivity calibration solution. If the value is accurate, it is not necessary to conduct 2 point calibration. If it is inaccurate, follow the steps in 3) to 4) to complete the 2 point of calibration using 12.88 mS/cm calibration solution.
Page 7
Symbol Self-Diagnosis information How to fix Make sure the probe is fully submerged in the solution. Check if the standard solution is expired or contaminated. Check if the conductivity electrode (two meter cannot recognize black rods) is damaged. Check if the conductivity electrode is conductivity standard solutions.
P ar am et e r S et ti ng Table of Settings Factory Symbol Parameter Setting Contents Content Default ˚C – ˚F ˚F Temperature Unit 5-20 seconds – Off Select automatic lock 1-8 minutes – Off Automatic Backlight Off 10-20 minutes –...
Page 9
Conductivity Calibration Reminder (P7) – set X hours (H) Or X days (D) in ZenTest mobile app – settings – Parameter – pH – Calibration Reminder. On the meter, you can only check the values that’s been set up on ZenTest App. For example, if 3 days is set up, the Er6 icon (see Figure-4) will appear in the lower right corner of the LCD screen in 3 days to remind you to perform calibration, also in the ZenTest App there will be a pop-up reminder.
Page 10
To replace a probe: 1)take off the probe cap; 2) screw off the probe ring 3) unplug the probe; 4) plug in the new replacement probe (pay attention to the probe’s position); 5) screw on the probe ring tightly. The replacement probe that is compatible with EC60-Z is: EC60-DE - 10 -...
Page 11
APERA INSTRUMENTS, LLC, any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS, LLC for a period of TWO YEARS (SIX MONTHS for the probe) from the delivery. This limited warranty does NOT cover any damages due to: accidental damage, unauthorized repair, normal wear and tear, or external causes such as accidents, abuse, or other actions or events beyond our reasonable control.