Apera Instruments 28APE190026 Instruction Manual

Premium conductivity tester (ec/tds/salinity/temp.)

Advertisement

Quick Links

EC60 Premium Conductivity Tester
(EC/TDS/Salinity/Temp.)
Instruction Manual
APERA INSTRUMENTS (Europe) GmbH
www.aperainst.de
V6.3

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 28APE190026 and is the answer not in the manual?

Questions and answers

Summary of Contents for Apera Instruments 28APE190026

  • Page 1 EC60 Premium Conductivity Tester (EC/TDS/Salinity/Temp.) Instruction Manual APERA INSTRUMENTS (Europe) GmbH www.aperainst.de V6.3...
  • Page 2: Table Of Contents

    Thank you for purchasing Apera Instruments EC60 Premium Conductivity Tester. Please carefully read this instruction manual before using the product in order to have an accurate and reliable test result, and avoid unnecessary damages to the meter or probe. For video tutorials, please go to www.aperainst.de Contents 1.
  • Page 3: Battery Installation

    1. Battery Installation Please install batteries according to the following steps. Please note direction of batteries: All POSITIVE SIDES (“+”) FACING UP. (Wrong installation of batteries will cause damage to the tester and potential hazards) ① Pull the battery cap up ②...
  • Page 4: Complete Kit

    3. Complete Kit Diagram - 2 4.Conductivity Calibration 4.1 Press key to switch to conductivity measurement mode. Rinse the probe in distilled water and dry it. 4.2 Pour certain amount (about half volume of the calibration bottle) of 1413μS/cm and 12.88 mS/cm conductivity calibration solution into accordant calibration bottles.
  • Page 5: Conductivity Measurement

    5. Conductivity Measurement Press key to turn on the tester. Rinse the probe in distilled water and dry it. Dip the probe in sample solution, stir gently, and allow it to stand still in the solution until a stable reading is reached. Get readings after comes up and stays.
  • Page 6: Parameter Setting

    5.4 TDS and conductivity is linear related, and its conversion factor is 0.40-1.00. Adjust the factor in parameter setting P5 based on the requirements in different industries. The factory default setting is 0.71. Salinity and conductivity is linear related, and its conversion factor is 0.5. The tester only needs to be calibrated in Conductivity mode, then after calibration of conductivity, the meter can switch from conductivity to TDS or salinity.
  • Page 7: Technical Specifications

    6.3 Parameter Setting Instruction a) Automatic lock (P1): Select “On” to activate auto lock function. When reading is stable for more than 10 seconds, the tester will lock the value automatically, and HOLD icon will display on LCD. Press key to cancel auto hold. b) Backlight (P2) “Off”-turn off backlight, “On”-turn on backlight, 1- backlight will last for 1 minute.
  • Page 8: Icons And Functions

    We warrant this instrument to be free from defects in material and workmanship and agree to repair or replace free of charge, at option of APERA INSTRUMENTS (Europe) GmbH, any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS (Europe) GmbH for a period of two years from the delivery (a six-month limited warranty applies to probes).