Apera Instruments PH60F-Z Instruction Manual

Apera Instruments PH60F-Z Instruction Manual

Smart ph tester
Hide thumbs Also See for PH60F-Z:

Advertisement

Quick Links

60-Z Series Smart pH Tester
(
pH/Temp./ORP)
Instruction Manual
PH60-Z pH Tester
PH60S-Z Spear pH Tester
PH60F-Z Flat pH Tester
www.aperainst.de
V1.0
- 1 -

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PH60F-Z and is the answer not in the manual?

Questions and answers

Summary of Contents for Apera Instruments PH60F-Z

  • Page 1 60-Z Series Smart pH Tester pH/Temp./ORP) Instruction Manual PH60-Z pH Tester PH60S-Z Spear pH Tester PH60F-Z Flat pH Tester www.aperainst.de V1.0 - 1 -...
  • Page 2: Table Of Contents

    Contents Introduction ......................- 3 - Battery Installation ....................- 4 - Keypad Functions ..................... - 4 - Complete Kit ......................- 5 - Things to Know Before Use ..................- 5 - pH Calibration ......................- 6 - pH Measurement ...................... - 8 - ORP Measurement....................
  • Page 3: Introduction

    1 Introduction Dear Customer, Thank you for choosing Apera Instruments PH60F-Z Smart pH Tester. Please carefully read this manual before using the product in order to have a reliable testing experience. 1.1 This product is designed with a two-way control on both the tester and ZenTest Mobile App.
  • Page 4: Battery Installation

    2 Battery Installation Please install batteries according to the following steps. *Please note direction of batteries: All POSITIVE SIDES (“+”) FACING UP. (Wrong installation of batteries will cause damage to the tester and potential hazards) ① Pull the battery cap up ② Slide the battery cap along to the direction of arrow ③...
  • Page 5: Complete Kit

    Distilled or deionized water (8-16oz) for rinsing the probe after each test b) Tissue paper for drying the probe 3) For PH60-Z and PH60F-Z testers: a) A few drops of water are added to the probe cap to keep the pH electrode in an activated state before the tester leaves factory.
  • Page 6: Ph Calibration

    6 pH Calibration 6.1 Calibration Steps 1) Short press to turn on the meter; rinse the probe in distilled water, shake the meter in the air and use tissue paper to dap off excess water (never rub or wipe the sensor tip). 2) Pour certain amount of pH 7.00 and pH 4.00 buffer solution into the corresponding calibration bottles (to about half volume of the bottle);...
  • Page 7 3) For pH Calibration buffer solutions, we recommend that users replace new buffer solution after 10 to 15 times of use to keep the standard buffer’s accuracy. Do NOT pour the used calibration solutions back into the solution bottles in case of contamination. 4) When calibrating strong basic solutions (pH>10), use 7.00 pH and 12.45 pH to calibrate;...
  • Page 8: Ph Measurement

    Solid samples such as cheese, meat, fruit, vegetables, sushi rice, sensor bread, and soil; also works well for general water solution. Flat surface measurement such as skin, paper, fabric, leather and PH60F-Z – Flat glass so on; micro sample testing; also works well for general water membrane solution.
  • Page 9: Orp Measurement

    3) PH60F-Z Flat probe testers are mostly for flat surface sample test. a) For skin test: skin should be without sweat or dirt, nor be overly cleaned (do not use facewash products before testing) to avoid affecting measurement results, dampen skin with some distilled water, slightly force flat probe onto the skin, get readings after value stabilized.
  • Page 10: Technical Specifications

    2) Auto. Lock (P02) – Users can set the auto lock time from 5 to 20 seconds. For example, if 10 seconds is set, when the measured value is stable for more than 10 seconds, the measured value will be automatically locked, and the HOLD icon will be displayed. Short press release the lock.
  • Page 11: Icons And Functions

    We warrant this instrument to be free from defects in material and workmanship and agree to repair or replace free of charge, at option of APERA INSTRUMENTS (Europe) GmbH, any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS (Europe) GmbH for a period of TWO YEARS (SIX MONTHS for the probe) from the delivery.

This manual is also suitable for:

Ph60-zPh60s-z60-z series

Table of Contents