Apera Instruments EC60 Instruction Manual

Apera Instruments EC60 Instruction Manual

Premium conductivity tester (ec/tds/salinity/temp.)
Hide thumbs Also See for EC60:

Advertisement

Quick Links

EC60 Premium Conductivity Tester
(EC/TDS/Salinity/Temp.)
Instruction Manual
APERA INSTRUMENTS, LLC
www.aperainst.com
1
V6.5

Advertisement

loading
Need help?

Need help?

Do you have a question about the EC60 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Apera Instruments EC60

  • Page 1 EC60 Premium Conductivity Tester (EC/TDS/Salinity/Temp.) Instruction Manual APERA INSTRUMENTS, LLC www.aperainst.com V6.5...
  • Page 2 Battery Installation Please install batteries according to the following steps. Please note direction of batteries: All POSITIVE SIDES (“+”) FACING UP. (Wrong installation of batteries will cause damage to the tester and potential hazards) ① Pull the battery cap up ②...
  • Page 3 Keypad Functions Short press — < 2 seconds Long press — > 2 seconds Short press to turn on the tester and long press to turn off the tester. When turned off, long press to enter parameter setting. In measurement mode, short press to turn on backlight.
  • Page 4 Conductivity Calibration 4.1 Press key to switch to conductivity measurement mode. Rinse the probe in distilled water and dry it. 4.2 Pour certain amount (about half volume of the calibration bottle) of 1413μS/cm and 12.88 mS/cm conductivity calibration solution into accordant calibration bottles.
  • Page 5 Conductivity Measurement Press key to turn on the tester. Rinse the probe in distilled water and dry it. Dip the probe in sample solution, stir gently, and allow it to stand still in the solution until a stable reading is reached. Get readings after comes up and stays.
  • Page 6 default setting is 0.71. Salinity and conductivity is linear related, and its conversion factor is 0.5. The tester only needs to be calibrated in Conductivity mode, then after calibration of conductivity, the meter can switch from conductivity to TDS or salinity. 5.5 Conversion Example: if conductivity measurement is 1000µS/cm , then the default TDS measurement will be 710 ppm (under the default 0.71 conversion factor), and the...
  • Page 7 Parameter Setting 1) Parameter Setting Chart Symbol Content Parameter Factory Default Auto Hold Off-On Backlight Off-1-On 1 (1 min auto-off) Temp. Compensation Coefficient 0.00 to 4.00% 2.00% TDS Coefficient 0.40 to 1.00 0.71 Salinity Unit ppt-mg/L ˚C-˚F ˚C Temperature Unit Restore to Factory Defualt No-Yes 2) Parameter Setting...
  • Page 8 Carrying case: 255*210*50mm/438g Probe Replacement Twist off the probe ring, unplug the probe, plug in the new replacement probe (pay attention to the electrode’s position), and screw on the probe ring. The model for EC60 tester’s replacement probe is EC60-E (AI209)
  • Page 9 We warrant this instrument to be free from defects in material and workmanship and agree to repair or replace free of charge, at option of APERA INSTRUMENTS, LLC, any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS, LLC for a period of two years from the delivery (a six-month limited warranty applies to probes).
  • Page 10 APERA INSTRUMENTS, LLC Address: 6656 Busch Blvd. Columbus, Ohio, 43229 Tel: 1-614-285-3080 Email: info@aperainst.com Website: www.aperainst.com...