1
Assembly for the dual-viewing (face-to-face) model
Clamp for 2
Clamp for 1
2
Assembly for the dual-viewing (side-to-side) model
1 Check that the serial numbers on the two components (main observer section and
secondary observer section) of the teaching head S-S are identical. (If the numbers are
different, observation will be adversely affected. In this case, please contact your
dealer or nearest Nikon representative.)
2 Adjust the height of the support stand so that the main observer section and the
secondary observer section are horizontal. This step is facilitated by first adjusting the
support stand to an approximate height before attaching the stand.
3 When a pointer is going to be used, assemble the system so that the secondary observer
will be positioned on the right. If reversed, the pointer control section will be at the
rear of the microscope.
4 When mounting the attachment on the microscopes ECLIPSE E800 and E1000, or on
the inverted microscopes ECLIPSE TE300 and TE200, separately available
components are required.
Mounting on the ECLIPSE E400, E600 or E600W
Clamp for 3
Pointer mounting
surface
Clamp for 2
2
1
Nikon
Y-THF
JAPAN
0100001
Clamp for 5
5
3
Main observer
2
section
Place
horizontally
III. Assembly
4
3
1
Check that the
serial numbers of
Y-THF
Nikon
the teaching head
0100001
S-S are the same.
7
Clamp for 7
Secondary
observer
6
section
Clamp for 6
4
Height adjustment
clamp screw
Adjust to an
approximate height
and readjust later.
JAPAN
Upper surface should
also be horizontal.