System Delivery Procedure
4. Adjust the ROI size: Press the Next key, then roll the trackball to adjust the ROI size.
5. Adjust the ROI steer: Press Next, then roll the trackball to steer the ROI.
6. Press the Select key to lock these settings.
Fig. 31:
7. Adjust the color gain: Rotate the C knob
Verify color fills the ROI area.
8. Press the C knob

Verify Doppler Mode

1. Press the PW knob
The system enables Doppler imaging. The Doppler tools and Doppler cursor become
active. The Doppler cursor changes to green when active.
2. Press the PW knob
3. Verify a Doppler strip displays on the system monitor.
It may be necessary to rotate the PW knob
4. Press the PW knob

Verify M Mode

1. Press the M knob
The system enables M Mode. The cursor becomes active. The cursor changes to green
when active.
2. Press the M knob
3. Verify a M-node strip displays on the system monitor.
It may be necessary to rotate the PW knob
4. Press the M knob
Siemens
NOTE
Roll the trackball to move position of the ROI color image.
(7/Fig. 11 / p. 19)
(9/Fig. 11 / p.
(9/Fig. 11 / p. 19)
(9/Fig. 11 / p. 19)
(1/Fig. 11 / p.
(1/Fig. 11 / p. 19)
(1/Fig. 11 / p. 19)
US15-101.812.01.02.02
05.10
(7/Fig. 11 / p.
again to exit Color Mode.
19).
again.
(9/Fig. 11 / p. 19)
again to exit Doppler Mode.
19).
again.
(9/Fig. 11 / p. 19)
again to exit Doppler Mode.
Page 29 of 40
Ultrasound
19).
clockwise to see the strip.
clockwise to see the strip.
ACUSON S2000
29
0
0

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents