Verify Transducer Initialization - Siemens ACUSON S2000 Installation Instructions Manual

Hide thumbs Also See for ACUSON S2000:
Table of Contents

Advertisement

20
1. Press the control panel Release button at the middle of the control panel handle.
Fig. 12: Control Panel Movement
2. Verify the control panel can be raised and lowered.
- Raise the control panel to its highest position.
- Lower the control panel to its lowest position
3. Verify that the control panel can be swiveled from left to right.
- Swivel the control panel to the left.
- Swivel back to the center locked position.
- Swivel the control panel to the right.
- Swivel back to the center locked position.

Verify Transducer Initialization

Use the following process to verify the transducers and the transducer ports:
1. Verify one of the transducers work in the left transducer port, see
ducer Port / p. 21)
2. Using the same transducer that was verified in the left port, verify the function of the
center transducer port, see
3. Using the same transducer that was verified in the center port, verify the function of the
right transducer port
4. Verify one at a time, each transducer shipped with the system in any of the verified
transducer ports.
5. Leave the last transducer being verified connected to the transducer port.
ACUSON S2000
for instructions.
(Verify the Center Transducer Port / p. 24)
(Verify the Right Transducer Port / p.
US15-101.812.01.02.02
05.10
System Delivery Procedure
26).
Page 20 of 40
Ultrasound
(Verify the Left Trans-
for instructions.