Keysight M9485A Startup Manual page 28

Pxie multiport vector network analyzer
Hide thumbs Also See for M9485A:
Table of Contents

Advertisement

Getting Started
STEP 3. Install the Network Analyzer Modules
b.
Figure 1-20 on page 28
Configurable Test Set measurement configuration. This is the
connection in both situations, where coupler M9378A is to the left side
of the M9377A and where M9378A is on the right side of the M9377A.
Figure 1-20 Cable Connections For Standard Measurement Configuration
c.
Figure 1-21 on page 28
power measurement. Its important to ensure that test port labeled 6
(ATT Out) is connected to test port 4 (Thru R). Similarly, connect test
port 7 (ATT In) to test port 5 (Thru T). This connection ensures that the
current flows through high power 20 dB attenuator.
Figure 1-21 Cable Connections For High Power Measurement Configuration
d.
Figure 1-22 on page 29
Dynamic Range measurement.
High Sensitivity measurement
High Gain measurement
below shows the cable connection for standard
below shows the cable connection for high
below shows the cable connection for Wide
28