7. Maintenance

7.1 General
1. Check main power cord and plug for any abrasion or excessive wear.
2. Check mattress cover for signs of wear or damage. Ensure mattress cover and tubes are connected
together correctly.
3. Check airflow from the quick connector. The airflow should alternate between each connector every
half-cycle time if it is in alternating mode.
4. Check the air hoses for kink or breaks. For replacement, please contact our local provider.

7.2 Fuse Replacement

1. Disconnect the plug from mains power when a blown fuse is suspected.
2. Use a proper tool to remove the cover of the fuse holder.
3. Insert a new fuse of the correct rating in, and replace the fuse holder back to correct position. The
fuse should be rated as T1.6AL /250V type and VDE approved.

7.3 Air Filter Replacement

1. Replace the air filter located at the back of pump.
2. The filter is reusable and can be washed gently with a mild detergent and water. Air dry the filter before
use.
3. Check and replace air filter regularly.