Digital Blood Pressure Monitor
Model UA-1020
Instruction Manual
Original
Manuel d'instructions
Traduction
Manual de Instrucciones
Traducción
Manuale di Istruzioni
Traduzione
使用手冊
翻譯
1WMPD4002224F

Advertisement

Table of Contents
loading

Summary of Contents for A&D UA-1020

  • Page 1 Digital Blood Pressure Monitor Model UA-1020 Instruction Manual Original Manuel d’instructions Traduction Manual de Instrucciones Traducción Manuale di Istruzioni Traduzione 使用手冊 翻譯 1WMPD4002224F...
  • Page 2: Table Of Contents

    Contents Dear Customers ..............2 Preliminary Remarks ............. 2 Precautions ................2 Contraindications............. 3 Parts Identification ..............4 Symbols ................5 Operation Mode ..............7 Using the Monitor ..............8 Installing / Changing the Batteries ........8 Connecting the Air Hose ..........9 Connecting the AC Adapter ..........
  • Page 3: Dear Customers

    Dear Customers Congratulations on purchasing a state-of-the-art A&D blood pressure monitor, one of the most advanced monitors available today. Designed for ease of use and accuracy, this device will facilitate your daily blood pressure regimen. We recommend that you read through this manual carefully before using the device for the first time.
  • Page 4: Contraindications

    removed from the electrical outlet when necessary. When reusing the device, confirm that the device is clean. Do not modify the device. It may cause accidents or damage to the device. To measure blood pressure, the arm must be squeezed by the cuff hard enough to temporarily stop blood flow through the artery.
  • Page 5: Parts Identification

    Parts Identification DC Jack Display Button Air Connector Plug Button Arm Cuff Button Button START Air Hose Air Socket Battery Compartment 1.5V Batteries Battery Cover (R6P, LR6 or AA) Display Average Pressure Settings MEMORY IHB/AFib symbol Cuff Fit Error Symbol Systolic Pressure Movement Error Symbol...
  • Page 6: Symbols

    Symbols Symbols that are printed on the device case Symbols Function / Meaning Standby and Turn the device on. Battery installation guide Direct current Serial number Date of manufacture Type BF: Device, cuff and tubing are designed to provide special protection against electrical shocks. EC directive medical device label WEEE label Manufacturer...
  • Page 7 Symbols Function / Meaning Recommended Action The reading may yield an Appears during measurement when incorrect value. Apply the the cuff is applied loosely. cuff correctly, and try the measurement again. Previous measurements stored in MEMORY. Average data FULL BATTERY The battery power indicator during measurement.
  • Page 8: Operation Mode

    Operation Mode 1. Normal Measurement Press the button. Blood pressure is measured and the data is stored START in memory. This device can store the last 90 measurements in memory. 2. Recalling the Data Press the button to recall the data in memory.
  • Page 9: Using The Monitor

    Using the Monitor Installing / Changing the Batteries Step 1 1. Remove the battery cover. Step 3 Step 1 2. Remove the used batteries and insert new batteries into the battery compartment as shown, Step 2 taking care that the polarities (+ and -) are correct.
  • Page 10: Connecting The Air Hose

    Using the Monitor Connecting the Air Hose Air socket Insert the air connector plug into the air socket firmly. Air connector plug Connecting the AC Adapter AC adapter plug Insert the AC adapter plug into the DC jack. Next, connect the AC adapter to an electrical outlet.
  • Page 11: Adjusting The Built-In Clock

    Using the Monitor Adjusting the Built-in Clock Press Adjust the clock prior to use. 1. Press the button until the year starts Year blinking. 2. Select the year using the button. Press the button to set the current year and move to month/day selection. The Month date can be set anywhere between the years 2010 and 2059.
  • Page 12: Selecting The Correct Cuff Size

    Using the Monitor Selecting the Correct Cuff Size Using the correct cuff size is important for an accurate reading. If the cuff is not the proper size, the reading may yield an incorrect blood pressure value. The arm size is printed on each cuff. The arm cuff is a consumable.
  • Page 13: How To Take Accurate Measurements

    Using the Monitor How to Take Accurate Measurements For the most accurate blood pressure measurement: Sit comfortably on a chair. Rest your arm on the table. Do not cross your legs. Keep your feet flat on the floor and straighten your back. Relax for about five to ten minutes before measurement.
  • Page 14: Measurements

    Measurements Before measurement, refer to “Notes for Accurate Measurement” on page 15. Normal Measurement At heart level Place the cuff on the arm (preferably the left arm). Sit quietly during measurement. Press the button. START Press All of the display segments are displayed. Zero (0) is displayed blinking briefly.
  • Page 15: Measurement With The Set Pressure

    Measurements The UA-1020 is designed to detect the pulse and to inflate the cuff to a systolic pressure level automatically. If re-inflation occurs repeatedly, use the following methods. Measurement with the SET Pressure During the blood pressure measurement, re-inflation may occur.
  • Page 16: Notes For Accurate Measurement

    Notes for Accurate Measurement Sit down in a comfortable position. Place your arm on a table with your palm facing upward and the cuff at the same level as your heart. Relax for about five to ten minutes before taking a measurement. If you are excited or depressed by emotional stress, the measurement will reflect this stress as a higher (or lower) than normal blood pressure reading and the pulse reading will usually be faster than normal.
  • Page 17 Measurement Using the TriCheck Mode Press the button. START All of the display segments are At heart level displayed. Zero (0) is displayed blinking briefly and the first measurement starts. Press When the measurement is complete, the systolic and diastolic pressure All of the display readings and pulse rate are displayed, segments displayed...
  • Page 18: Recalling The Memory Data

    Recalling the Memory Data Note: This device stores the last 90 measurements in memory. 1. Press the button. Press The average of all measurements and the number of data are displayed. Average systolic (If no data, “0” is displayed. Press the Average diastolic button to turn the START...
  • Page 19: What Is The Ihb/Afib Indicator

    What Is The IHB/AFib Indicator? When the monitor detects an irregular rhythm during the measurements, the IHB/AFib indicator will appear on the display with the measurement values. Note: We recommend contacting your physician if you see this IHB/AFib indicator frequently. What Is The AFib? The heart contracts due to electrical signals occurring in heart and sends blood through the body.
  • Page 20: About Blood Pressure

    About Blood Pressure What is Blood Pressure? Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts. Diastolic pressure occurs when the heart expands. Blood pressure is measured in millimeters of mercury (mmHg).
  • Page 21: Troubleshooting

    Take measurements at the same time every day using the procedure described in this manual to get to know your normal blood pressure. Regular readings give a more comprehensive blood pressure history. Be sure to note the date and time when recording your blood pressure.
  • Page 22: Maintenance

    However it is generally recommended to have the device inspected every 2 years, to ensure proper functioning and accuracy. Please contact the authorized dealer in your area or A&D for maintenance. Technical Data Type UA-1020 Measurement method Oscillometric measurement Measurement range Pressure: 0 - 299 mmHg...
  • Page 23 Weight Approx. 285 g, excluding the batteries Ingress protection Device: IP20 Applied part Cuff Type BF Useful life Device: 5 years (when used six times a day) Cuff: 2 years (when used six times a day) AC adapter: 5 years (when used six times a day) Accessory AC adapter The adapter is to connect the blood pressure monitor to a power source at home.
  • Page 24 A&D Company, Ltd. 1-243 Asahi , Kitamoto-shi, Saitama 364-8585, JAPAN Telephone: [81] (48) 593-1111 Fax: [81] (48) 593-1119 Emergo Europe B.V. Prinsessegracht 20, 2514 AP The Hague, The Netherlands Tel: [31] (70) 345-8570 Fax: [31] (70) 346-7299 A&D INSTRUMENTS LIMITED Unit 24/26 Blacklands Way, Abingdon Business Park, Abingdon, Oxfordshire OX14 1DY United Kingdom Telephone: [44] (1235) 550420...