Remove and replace battery if monitor is not used for more than six months. PRECAUTIONS The UA-704 is designed to be used at home, by those who are eighteen (18) years and older, to monitor blood pressure (systolic and diastolic) and pulse rate. It is not designed for...
Page 3
Congratulations on purchasing a state-of-the-art LifeSource blood pressure monitor—one of the most technologically advanced monitors available today. Designed for ease of use and accuracy, this monitor will facilitate your daily blood pressure regimen. Physicians agree that daily self-monitoring of blood pressure is an important step individuals can take to maintain cardiovascular health and prevent the serious consequences associated with undetected and untreated...
WHAT DISPLAY SYMBOLS MEAN Display Condition/Cause Recommended Symbol Action Appears during Remain still and refrain measurement; blinks when from talking; measurement MEASURING pulse is detected. in progress. Appears on display prior to Press exhaust button to and during measurement. release air. Blinks if there is air in cuff DEFLATE at power up.
HOW UA-704 WORKS LifeSource™ Blood Pressure Monitors are easy to use, accurate and digitally display full measurement readouts. Our technology is based on the “oscillometric method” — a noninvasive blood pressure determination. The term “oscillation” refers to any measure of vibrations caused by the arterial pulse.
11.8” - 17.7” (30-45cm) Large NOTE: The LifeSource small cuff is not compatible with the UA-704 monitor. Cuff Size Indicator On The Cuff Our cuff has an indicator which tells you whether you are using the correct cuff size or not. Place the cuff on your arm (see Page E-6 to learn how to put the cuff on correctly) and if ▲...
TAKING YOUR BLOOD PRESSURE Tips for Blood Pressure Monitoring: ☞ Relax for about 5 to 10 minutes before measurement. ☞ Remove constricting clothing and place cuff on bare arm. ☞ Unless your physician recommends otherwise, use left arm to measure pressure. Now you are ready.
Page 10
TAKING YOUR BLOOD PRESSURE ☞ IMPORTANT: Measure pressure at the same time each day. Press the START button Watch for all display symbols to appear briefly, followed by a flashing “0”, indicating that monitor is ready for measurement. If the mark is displayed, the cuff has some air trapped in it.
Page 11
TAKING YOUR BLOOD PRESSURE When the correct pressurization has been reached, the automatic exhaust mechanism will gradually reduce the pressure in the cuff. Sit quietly during measurement. symbol will blink with your pulse beat. CURRENT CUFF PRESSURE NOTE: Exhaust rate cannot be adjusted during measurement. If you wish to stop inflation at any time, press the exhaust valve to release the pressure in the cuff.
NOTE: Allow at least three minutes between measurements on the same person. WHAT IS AN IRREGULAR HEARTBEAT? The UA-704 Blood Pressure Monitor provides a blood pressure and pulse rate measurement even when an irregular heartbeat occurs. An irregular heartbeat is defined as a heartbeat that varies by 25% from the average of all heartbeat intervals during the blood pressure measurement.
ABOUT BLOOD PRESSURE ❤ What Is Blood Pressure? Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts; diastolic pressure occurs when the heart expands. Blood pressure is measured in millimeters of mercury (mmHg).
ABOUT BLOOD PRESSURE ❤ Assessing High Blood Pressure The following standards for assessing high blood pressure (without regard to age) have been established as a guideline. BP Classification Systolic (mmHg) Diastolic (mmHg) Normal <120 <80 Prehypertension 120-139 80-89 Stage 1 Hypertension 140-159 90-99 Stage 2 Hypertension...
❤ ABOUT BLOOD PRESSURE Why Measure Blood Pressure at Home? It is now well known that, for many individuals, blood pressure readings taken in a doctor’ s office or hospital setting might be elevated as a result of apprehension and anxiety.
Page 16
❤ In Canada (Toll-Free): 1-800-463-5414 E-13...
✓ TROUBLESHOOTING Corrective action Probable cause Problem Replace with a new battery. Display is blank, Battery is drained. even when the Place the battery with power is turned Battery terminals are negative and positive not in the correct terminals matching those position.
CONTACT INFORMATION Our products are designed and manufactured using the latest scientific and technological methods, and offer accurate, easy to use, home monitoring and treatment options. Our full product line includes: Aneroid Blood Pressure Kits Blood Pressure Cuffs Digital Blood Pressure Monitors Digital Thermometers Heart Rate Monitors Nebulizers...
Page 20
CONTACT INFORMATION IMPORTANT! If You Need Assistance with the Set-Up or Operation We Can Help! Please call us FIRST before contacting your retailer at LifeSource Health Line 1-888-726-9966 - Toll Free 1-800-463-5414 - In Canada A specially trained representative will assist you A division of A&D Engineering, Inc.