ERROR CODES

Following is a description of typical error codes that may appear on VI/VIR-Series balances.
1)
L
This indicates that the balance is under loaded. Check to see if the weighing
pan is properly installed and seated on the top of the unit.
2)
H
This indicates that there may be an excessive weight applied to the weighing pan. It is critical
not to exceed the maximum load capacity by more than 150% or damage to the balance may
occur.
3)
E-01
Capacity of LCD digit readout has been exceeded in CT and DWT modes.
4)
E-02
Usually detected when the calibration operation is being attempted. It is usually the result of
the balance being exposed to vibrations, air drafts, or unstable environments. The balance
must be placed in a stable environment before calibration occurs. Before pushing the
CAL/MODE button as part of the calibration process, the LCD should display a stable ZERO
with the mode arrow stable on the gram (g) mode.
5)
E-10
This will be displayed only if there is a value stored in memory and the tare button is pressed.
Memory must be cleared or the unit turned off in order to restore the normal tare function.
6)
E-30
This is an indication that an electronic malfunction has occurred. The unit must be returned to
the factory for evaluation.
7)
E-54
This fault will appear on the display when the unit is powered on. This occurs when the
electronics are no longer within the factory set parameters. The most common cause is from
objects being dropped onto the weighing pan or from the balance itself being dropped. The
balance must be returned to Acculab or an authorized service center for electronic evaluation
and repair.
L
8)
(in the upper left corner of display) Following are the most probable causes of this error code.
The balance must be returned to Acculab or an authorized service center for electronic
evaluation and repair.
NOTE: Do not attempt to access internal components or make any unauthorized repairs since this will void the
warranty.
SUPPLIED ACCESSORIES
• AC Adapter (DC 12V):
#AC-1 -For use in USA/Canada (AC 120V 60Hz)
#AC-2 -For use in Europe (AC 240V 50Hz)
• Calibration Weight:
#VI-1mg - One (1) 100g weight to calibrate balance at 100g OIML Class F1
#VI-3mg - Two (2) 100g weight to calibrate balance at 200g OIML Class F1
• Draft Shield & Spill Ring:
Glass Draft Shield with stainless steel cover
Stainless steel spill ring
OPTIONAL ACCESSORIES
• Cable lock, carry case
page 6
1. There is something causing the sensor to function out of specification.
2. Rough, improper handling of the balance.
3. Liquids and other materials spilled into the balance.
#AC-3 -For use in U.K. (AC 240V 50Hz)
#AC-4 -For use in Japan (AC 100V 50/60Hz)
#AC-5 -For use in AUS/NZ (AC 240V 50Hz)
Model #