72

Gas connections for Magnos28 (Model EL3040)

Magnos28: Gas connections
1
Sample gas inlet
Sample gas outlet
2
not assigned
3
not assigned
4
Purge gas inlet for case
11
12
Purge gas outlet for case
Pressure sensor
13
Design:
⅛ NPT female thread (stainless steel 1.4305/SAE 303)
Connection of flexible tubes: straight screwed fittings (PP) with nozzles for flexible tubes with
inside diameter = 4 mm (supplied)
Connection of pipelines: Threaded connections (not supplied)
Note:
The pressure sensor is installed as an option. It is connected to the port 13 via an FPM hose. The
pressure sensor connection and the sample gas outlet must be connected to each other via a
T-piece and short lines to ensure a precise pressure correction. The lines must be as short as
possible or if this is not practicable, the line must have a sufficiently wide internal diameter
(at least 10 mm) so that the flow effect is minimized. If the pressure sensor connection is not
connected to the sample gas outlet, the pressure sensor and the sample gas outlet must be on
the same pressure level.
EL3000 CONTINUOUS GAS ANALYZERS
|