Set the IP address in the gas analyzer
EL3000 CONTINUOUS GAS ANALYZERS
Setup
Device settings
Parameters
The parameters which have to be input depend on the DHCP setting:
DHCP on:
Network name (max. 20 characters, no blanks or special char-
acters),
DHCP off:
IP address, IP address mask and IP gateway address.
The network name can only be altered in the configurator. The default net-
work name consists of "EL3K" and the last six positions of the MAC address
(example: "EL3KFF579A").
If the parameter "DHCP" is set to "off", the Ethernet configuration is reset to
the standard configuration (default IP address); this will prevent the inad-
vertent assignment of an IP address from a DHCP pool.
Addresses
The IP address, IP address mask and IP gateway address must be obtained
from the system administrator.
NOTE
The address bits that can be varied in the address mask may not all be set to
0 or 1 (broadcast addresses).
MAC Address
The 12-digit MAC address is unique and stored in each device during manu-
facture. It cannot be altered.
|
CI/EL3000-EN REV. B
Ethernet
113