90
Maximum length of the sample gas line
Provide for sample gas line purging
EL3000 CONTINUOUS GAS ANALYZERS
Connection of the sample gas line
Heated sample gas line (tube with inside/outside diameter 4/6 mm)
1
O-ring 6.02 x 2.62
2
3
Socket
4
O-ring 12.42 x 1.78
Connection of a further sample gas line (tube with outside diameter
6 mm):
either with
5
Male fitting G¼"
or with
6
Socket
Tapered ring
7
Compression fitting
8
Closure:
Screw cap
9
O-ring 12.42 x 1.78
10
Waste gas tube
11
Sample gas
Connection of the sample gas line
inlets:
on 19-inch housing:
I
from the rear
II
from above
III
from below
The maximum length of the heated sample gas line (inside diameter 4 mm)
is 30 m.
Install a shut-off valve in the sample gas line (definitely recommended with
pressurized sample gas), in order to provide a means for purging the gas
line system by feeding in an inert gas, e.g. nitrogen, from the gas sampling
point.
|
CI/EL3000-EN REV. B
on wall-mount housing:
from below
from the right
not possible – the sample gas inlet
must always be closed