Operation: Advanced
42
Pump Diagnostics
To check the pumping mechanism functionality,enter the diagnostics menu. Press [4]
and press [4] again to advance to the next menu of tests. Press [1] to perform the
pump test. The display should read "Pump Test". Once the test has completed suc-
cessfully, the display will read Piston Test Success and return to the diagnostic menu.
Step*
Valve Diagnostics
To check the respiration valve's functionality, enter the diagnostics menu. Press [4]
and press [4] again to advance to the next menu of tests. Press [2] to perform the
valve test. The display should read "Valve Test". Once the test has completed suc-
cessfully, the display will read Valve Test Success and return to the diagnostic menu.
Step*