Diagnostics - Harvard Apparatus Inspira ASV MA1 55-7058 User Manual

Advanced safety ventilator
Table of Contents

Advertisement

Operation: Advanced
39

Diagnostics

To enter the Inspira diagnostic menu, first enter the hardware configuration menu and
press [4] to advance to the next menu. The display should read "Press 5= Diagnostics".
Press [5] to enter the diagnostics menu. The display should read "Press 1= Dsply,
2=Key, 3=Alarm, 4=Next".
Step*
Display Diagnostics
To perform the display diagnostic, enter the diagnostic menu. Press [1] to perform
the display test. The display should read "Display Test", then illuminate all pixels and
return to the diagnostics menu.
Step †
Keypad Diagnostics
To run the keypad diagnostic, enter the diagnostics menu. Press [2] to start the key-
pad diagnostic. The display should momentarily read "Keypad Test,Press Enter to Exit".
Then the display will show a cursor in its upper left-hand corner. This test echoes key-
strokes on the display. Press each key and verify that the display echo matches the key
pressed. Press [Enter] to exit this test and return to the diagnostics menu.
Step †
* Note: See page 37 for steps 1 through 3.
† Note: See above for steps 4 and 5.
Publication 5404-001-REV-C
Key
Key
Key
Display
Display
Display