Operation: Advanced
35
Parity Configuration
To configure the port parity settings, enter the port configuration menu. Press [3] to
enter the parity configuration menu. The display should read "Define Parity: 1= Even,
2= Odd, 3= None". Choose the desired parity by pressing the corresponding numer-
ic key. After making your selection Inspira will automatically return to the main
screen. The default parity is "none".
Step*
Address Configuration (Future Feature)
To assign an address to the Inspira for use in a daisy chain, enter the port setup menu.
Press [4] to enter the address configuration menu. The display should read "Enter
Address" and display Inspira's current address. Use the numeric keys to enter an
address between 0 and 999. Press [Enter] to enter the selected address and to return
to the main screen.
Step*
Testing RS232 Communication
Once Inspira and the external device are configured with the same port settings, con-
nect the Inspira to the external device using the proper cable. Send the VER command
from the external device to Inspira. If the RS232 connection is working properly the
Inspira will return its version number. If nothing is returned, check the cable con-
nection and the port settings to verify that they are configured correctly.
* Note: See page 33 for steps 1 through 4.
Publication 5404-001-REV-C