Operation: Getting Started
13
Figure 9. Main Operating Screen, Volume-Control Mode
Turning Ventilator On
Power must be properly supplied to the ventilator, according to the power specifica-
tions. Locate the power switch on the rear of the ventilator. Toggle the switch so that
the "I" side of the switch is pressed down. The ventilator is now powered on.
Volume-Controlled Ventilation
With machine powered ON, press and release the SELECT MODE key until VOL mode
is selected. Press SET and WEIGHT and enter the animal's weight in kg. Press and hold
parameter keys in order to verify respiration parameters. If any parameters need to be
changed, press the SET key and the appropriate parameter key and enter the value to
set. Press and release the RUN/STOP key to start ventilation.
Pressure-Controlled Ventilation (Inspira ASV p Only)
With machine powered ON, press and release the SELECT MODE key until PRES mode
is selected. Press SET and WEIGHT and enter the animal's weight in kg. Press and hold
parameter keys in order to verify respiration parameters. If any parameters need to be
changed, press the SET key and the appropriate parameter key and enter the value to
set. Press and release the RUN/STOP key to start ventilation.
Publication 5404-001-REV-C