Operation: Getting Started
16
Safe Range
Setting up the ASV ventilator is both easy and safe thanks to the Safe Range
Once the animal's body weight has been entered the software computes the median
settings for tidal volume and respiratory rate.These same equations are used to cal-
culate a Safe Range
calculate the correct tidal volume and respiratory rate for the animal. The tidal vol-
ume (in liters) is determined by the equation 0.0062 x M b
mass in kilograms.
The respiratory rate (in min
Range
or equal to ±10% of the calculated tidal volume and respiratory rate. All the default
parameter settings can be overridden. If you enter a setting outside of the Safe Range
for the body mass of the animal to be ventilated you are warned that the value is out-
side the Safe Range™ and asked to confirm the setting.
SAFE RANGE – Pressure-Control Mode
Inspiration Pressure Range
Safe Range: 15 to 25 cmH 2 O
Full Range: 5 to 50 cmH 2 O
Sigh Pressure Range
Safe Range: Insp. Pressure to (Insp. Pressure + 10cmH 2 O) or 35cmH 2 O,
whichever is less
Full Range: Insp. Pressure to 50cmH 2 O
Publication 5404-001-REV-C
Respiratory Profiles
around the median values. Inspira uses the entered weight to
-1
is established by computing a range of acceptable values, which are less than
) is determined by the equation 53.5 x M b
software.
1.01
where M b is the animal
-0.26
. The Safe