Operation: Advanced
38
I/O Setup
To review or modify the I/O settings, enter the hardware configuration menu. Press [2]
to enter the I/O setup. The display should now read "Set I/O: Press 1= Digital, 2=
Analog". Pressing [1] selects Digital I/O configuration. This feature is a future option
and is not presently available. Pressing [2] selects the Analog I/O menu. The display
should now read "Analog 1= Arwy Pres". This indicates that the airway pressure ana-
log signal is connected to Analog Out 1. This is the only analog signal available in this
model. Inspira then returns to the main display screen.
Step*
Display Brightness Level
To change the display brightness level, enter the hardware configuration menu. Press
[3] to enter the display configuration. The display should now read "Set Brightness
Level 100%" (default setting). There are four options for the display brightness, 25, 50,
75, and 100%. Use the [Select Mode] key to toggle through the different levels of
brightness. The [enter] key selects the brightness level and then returns to the main
screen.
Step*
* Note: See page 37 for steps 1 through 3.